U2802A
Table 3. Measurement accuracy of the U2356A at 0 to 18 °C and 28 to 45 °C
Thermocouple measurement accuracy (U2356A @ 0 to 18 °C and 28 to 45 °C)
T/C type
ITS-90 Temperature range (°C)
Optimum measurement range
(°C)
Without
averaging
(± °C)
50 points
averaging
(± °C)
500 points
averaging
(± °C)
Low High Low High
B 0 1820 1100 1820 6.1 3.1 2.4
400 1100 14.4 6.3 2.7
E –270 1000 –150 1000 3.0 2.6 2.6
–200 –150 4.2 3.7 3.6
J –210 1200 –150 1200 2.9 2.5 2.5
–210 –150 4.9 4.1 4.0
K –270 1372 –100 1200 3.3 2.9 2.9
–200 –100 5.3 4.2 4.0
N –270 1300 –100 1300 3.4 2.7 2.6
–200 –100 6.8 4.6 4.1
R –50 1768 300 1760 6.2 3.7 3.2
–50 300 15.7 7.2 3.8
S –50 1768 400 1760 6.4 4.0 3.4
–50 400 14.2 6.6 3.4
T –270 400 –100 400 3.0 2.4 2.2
–200 –100 5.3 4.2 3.9