Agilent Technologies Sprinkler 54753a Instrukcja Użytkownika Strona 38

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 44
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 37
Command
:MEASure:VTOP[?] [<source>]
Measures the statistical top of the selected source waveform.
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Command
:MEASure:VUPper[?] [<source>]
Measures the voltage value at the upper threshold of the waveform.]
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Measure Commands
36
Przeglądanie stron 37
1 2 ... 33 34 35 36 37 38 39 40 41 42 43 44

Komentarze do niniejszej Instrukcji

Brak uwag