Agilent Technologies Sprinkler 54753a Instrukcja Użytkownika Strona 36

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 44
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 35
Command
:MEASure:VMIN[?] [<source>]
Measures the absolute minimum voltage present on the selected source
waveform. When FFT is the specified source, the minimum value in the
spectrum is measured.
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number> | FFT}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Command
:MEASure:VPP[?] [<source>]
Measures the absolute maximum and minimum voltages on the selected
source, then calculates the peak-to-peak voltage as the difference between
the two voltages. When FFT is the specified source, the range of values in the
spectrum is measured.
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number> | FFT}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Measure Commands