Command
:MEASure:NWIDTh[?] [<source>]
Measures the width of the first negative pulse on the screen using the
mid-threshold levels of the waveform (50% levels with standard
measurements selected).
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Command
:MEASure:OVERshoot[?] [<source>]
Measures the width of the overshoot of the first edge on the screen.
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Command
:MEASure:PERiod[?] [<source>]
Measures the period of the first complete cycle on the screen using the
mid-threshold levels of the waveform (50% levels with standard
measurements selected).
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}