50 N2820/1A User’s Guide
2 Probing
To Add an R
SENSE
Resistor to an N2825A Head
Figure 28 Location of R
SENSE
Resistor (R1)
4 Replace the head’s cover using the four screws. (M1.6 x 0.35,
5 mm long, Agilent part number 0515- 5210)
CAUTION To avoid damaging the aluminum casting, do not over tighten the screws.
5 Connect the N2825A to an N2820/1A probe and connect the
probe to the oscilloscope.
6 On Infiniium oscilloscopes,
a Click the Setup > Probe Configuration menu command.
b Select the tab for the probe’s primary channel. N2821A
probes only have a primary channel.
c In the User Def Head Resistance field, enter the resistance of
the DUT’s R
SENSE
resistor in ohms.
d In the dialog box, click Add Head to give your head with
associated external R
SENSE
resistor a unique name.
7 On InfiniiVision oscilloscopes,
a Press the front- panel channel key that is associated with
the probe’s primary input.
b Click the Probe softkey.
c Press the R-Sense softkey and enter the resistance of the
R
SENSE
resistor that you are using.