Agilent Technologies N1022A Podręcznik Użytkownika Strona 19

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 82
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 18
General Information 1
N2820A Probes and Dual-Grid View
N2820/1A User’s Guide 19
If you’re familiar with the oscilloscope’s Zoom Mode, you
may be interested to know that while Zoom Mode uses
software to expand the zoomed- in waveform, dual- grid view
uses hardware amplification applied by the probe’s high- gain
amplifier.
On the display, the white bounding- area box identifies the
portion of the zoomed- out waveform that is displayed in the
zoomed- in window. This box does not indicate waveform
clipping.
Dual- grid view allows simultaneous viewing of high
sensitivity and high dynamic- range inputs. One example of
using dual- grid view is to quickly identify and view a
specific pulse and perform a DC power measurement on that
pulse.
To view only one window, click Setup > Display to open the
Display Setup dialog box and, in the Quantity field, select 1.
NOTE For best measurement results, on the oscilloscope click Setup >
Acquisition and select one of the High Resolution settings.
Przeglądanie stron 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 81 82

Komentarze do niniejszej Instrukcji

Brak uwag