General Information 1
Connecting the Probe to The Oscilloscope
N2820/1A User’s Guide 17
To configure the probe
1 Connect the probe to the oscilloscope. On N2820A probes,
connect the primary channel first followed by the secondary
channel.
2 Press the front- panel channel key that is associated with the
probe’s primary input.
3 Press the Probe softkey.
4 If you are using an N2821A probe, press the Zoom-In softkey
to toggle between zoomed- in (blue selection box on softkey)
or zoomed- out (clear selection box) for the channel.
5 If you are using an N2825A user- defined head, press the
R-Sense softkey, shown in Figure 8 on page 16, and enter the
resistance of the R
SENSE
resistor that you are using.