Agilent Technologies Ininiium 90000 Dokumentacja Strona 37

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 36
Chapter 3: Testing Performance
Analog Bandwidth - Maximum Frequency Check
37
c Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
d Pull down the Setup menu, select Acquisition and then set up the acquisition parameters
as follows:
Memory Depth = Automatic
Sampling rate = Maximum (40 GSa/s)
Sin(x)/x Interpolation filter enabled
Averaging = Disabled
e Pull down the Measure menu, select Voltage and then select V rms.
Click here and enter 16E-9
Przeglądanie stron 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 147 148

Komentarze do niniejszej Instrukcji

Brak uwag