Chapter 3: Testing Performance
DC Gain Accuracy Test
32
c Select the Vavg measurement as shown below.
d When the Enter Measurement Info window is displayed, ensure that the V avg function
is set up as follows and then click OK:
Source = Channel 1
Measurement Area = Entire Display
5
Press the Clear Display key on the scope, wait for the #Avgs value (top left corner of
screen) to return to 256 and then record the scope's mean V avg reading in the DC Gain
Test section of the Performance Test Record.
Notes
For all scope readings in this procedure, use the mean value in the Measurements display area at the
bottom of the screen.
If a question mark is displayed in front of any of the values at the bottom of the screen, press the Clear
Display key on the scope, wait for the #Avgs value to return to 256 and then record the scope reading.
V avg
measurement