Przeglądanie stron 92
16. Harmonic and Intermodulation Distortion Test
18.
Press
@ZZi%J
on the analyzer. Key in the following:
&ENTER
FREQUENCY_)
......................................
7.2
GHz
CFREQUENCY
SPAN)
.........................................
100
kHz
19.
20.
Press MARKER (PEAK SEARCH),
(GiCCXF].
Set (FREQUENCY
SPAN)
to 10
kHz.
Press MARKER [PEAK SEARCH),
I-),
[
MKR/A+
STP SIZE .
21.
Adjust synthesized sweeper output level to place peak of signal
trace at the reference level line.
22.
23.
24.
Press DISPLAY LINE
CENTER]
and key in -80
dBm.
Activate (CENTER FREQUENCY) and press
@)
to tune to the second
harmonic of the input signal.
Key in the following:
(REFERENCE LEVEL]
.......................................
-20
dBm
[FREQUENCY SPAN]
...........................................
1
kHz
25.
The second harmonic should be below the display line
(<-
100
dBc).
Intermodulation 26.
Distortion
Second harmonic level of 7200 MHz
dBc
Set both synthesized sources as follows:
RANGE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . + 10 dBm
METER MODE
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LEVEL
RF OUTPUT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
ALC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . INT
AM
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
FM DEVIATION MHz . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . OFF
27. Connect equipment as shown in Figure 2-29 with “output”
of power divider connected to power sensor. The FREQ
REFERENCE switch on the rear panel of the analyzer should be
set to INT and the FREQ REFERENCE switch on both synthesized
source rear panels should be set to EXT.
28. Set one synthesized source for an output frequency of 2099.500
MHz, the other to 2100.500 MHz.
29. Set one synthesized source RF OUTPUT switch to ON and adjust
the output power level for a power meter indication of -25.00
dBm
f0.20
dB.
Return the RF OUTPUT switch to the OFF setting.
2-66