Agilent Technologies 8920A
RF Communications Test Set
Programmers Guide
Firmware Version A.18.00 and above
Agilent Part No. 08920-90220
Printed in U. S. A.
April 2000
Rev. B
DUPLETXRX PREV TESTS
CONFIHELPMSSG HOLD PRINT
SCREEN CONTROL
LOCAL
ADRS
RECAL
SAVE
MEAS
PRESE
INSTRUMENT STATE
DATA FUNCTIONS
INCR
REF
INCR
METER
INCR
AV G
LO HI
CURSOR CON-
PUSH TO
CANCE
SHIFT
k1
k1’
k2
k2’
k3
k3’
k4
ASSIG
k5
RELEA
USER
DATA
789
456
123
0
+
_
ENTER
GHz
dB
MHz
%
kHz
s
Hz
ms
%
ppm
NO
ON/OFF
YES
MEMO
AUDIO IN
LOHI
!
MAX
!
MAX
AUDIO SQUELCVOL-
MIC/
MAX PO WER 200
!
ANT INDUPLEX OUTRF IN/OUT
!
MAX POWER
POWE
OF O
Przeglądanie stron 0
1 2 3 4 5 6 ... 580 581

Podsumowanie treści

Strona 1 - Programmer’s Guide

1Agilent Technologies 8920A RF Communications Test SetProgrammer’s GuideFirmware Version A.18.00 and aboveAgilent Part No. 08920-90220Printed in U.

Strona 2

10 Service and SupportAny adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer

Strona 3

100S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg1.sec AF Generator 1AF Generator 1 1In setting AFGenerator 1, you must first select a destination

Strona 4

101AF Generator 2 Pre-Modulation FiltersAF Generator 2 Pre-Modulation Filters To improve performance, one of four pre-modulation filters is automatic

Strona 5

102S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/EncoderAF Generator 2/Encoder ’spaceSingleBurstContStepReturns quoted str

Strona 6 - Warranty: 1 year

103AF Generator 2/Encoder:AMPS or :TACS1 AM, FM, and LEVel correspond to the setting of the AFGen2 To field.’spaceIldeBusyWS Delay1stBitDlyReturns qu

Strona 7

104S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder1 AM, FM, and LEVel correspond to the setting of the AFGen2 To fie

Strona 8 - Manufacturer’s Address:

105AF Generator 2/Encoder:NAMPs or :NTACs’spaceNAMPSNTACSReturns quoted string?’:AFGenerator2:STANDard:ENCoder’spaceIldeBusyWS Delay1stBitDlyReturns

Strona 9

106S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder:AFGenerator2:ENCoder:AFG2 continued:NAMPs:NTACs:DSAT:STOP:SEND:ME

Strona 10 - Service and

107AF Generator 2/Encoder:CDCSs and :DTMF:AFGenerator2:ENCoder:AFG2 continued:CDCSs:CODE’spacestringReturns quoted string?’:RATESee Real Number Setti

Strona 11 - Manufacturer’s

108S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder:DPAGing:AFGenerator2:ENCoder:DPAGing:GSC:CODE’spacestringReturns

Strona 12

109AF Generator 2/Encoder’spaceGSCPOCSAGReturns quoted string?’:AFGenerator2:STANdard:ENCoder:AFG2 continued:DPAGing:RATESee Real Number Setting Synt

Strona 13 - Contents

11Manufacturer’s DeclarationThis statement is provided to comply with the requirements of the German Sound Emission Directive, from 18 January 1991.

Strona 14

110S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder:FGENerator and :TSEQuential:AFGenerator2:ENCoder:AFG2 continued:F

Strona 15

111AF Generator 2/Encoder:NMT:AFGenerator2:ENCoder:NMT :AINformation’spacestringReturns quoted string?’(Add Info):BSIDentity:MSNumber:MAINtenance:PAS

Strona 16

112S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder:AFGenerator2:ENCoder:NMT ’spaceMSBSMTXReturns quoted string?’:DUT

Strona 17

113AF Generator 2/Encoder:LTR:AFGenerator2:ENCoder:LTR:AREA1:AREA2:FREE1:FREE2:GOTO1:GOTO2:HOME1See Integer Number Setting Syntax:HOME2:ID1:ID2:AFG2

Strona 18

114S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder:EDACs:AFGenerator2:ENCoder:AFG2 continued:EDACs:CNCHSee Integer N

Strona 19

115AF Generator 2/Encoder:MPT1327:AFGenerator2:ENCoder:MPT1327:SIDentity(System Identity)(Prefix):PREFix:SCU:RUUT See Integer Number Setting SyntaxSe

Strona 20

116S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afg2_enc.sec AF Generator 2/Encoder1 Integer value from 1 to 322 String maximum length 300:AFGenerato

Strona 21

117Configure, I/O ConfigureConfigure, I/O Configure NOTE: The CONFIGURE screen RF Display, RF Chan Std, User Def Base Freq, Chan Space, and(Gen)-(Anl

Strona 22

118S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\conf.sec Configure, I/O Configure:CONFigure’spaceOnOffReturns quoted string?’:KNOB’space1 min2 min5 m

Strona 23

119Configure, I/O Configure:CONFigure:OFLevel(RF Level Offset)(RFGen Volts)’spaceOnOffReturns quoted string?’:MODE:CONF continuedSee Real Number Sett

Strona 24

12 Chapter 7, IBASIC Controller, describes how to develop Instrument BASIC (IBASIC) programs for use on the Test Set’s built-in IBASIC Controller. To

Strona 25 - Using GPIB

120S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\conf.sec Configure, I/O Configure:CONFigure:SPB(Port B)’space150300600120024004800960019200Returns qu

Strona 26 - Overview of the Test Set

121Configure, I/O Configure:CONFigure’spaceOnOffReturns quoted string?’:IBECho:SPORt(Port A)’space150300600120024004800960019200Returns quoted string

Strona 27 - Manual Control Mode

122 Call ProcessingCall Processing’spaceMEASDATAReturns quoted string?’:CALLP:CPRocess:MODE:ACTive’spaceAMPSTACSJTACSReturns quoted string?’:CSYStem:

Strona 28

123Call Processing’space10 characters maxReturns quoted string?’:PNUMber’space5970Hz6000Hz6030HzReturns quoted string?’:SATone’spacePHONE NUMMIN2 MIN

Strona 29

124 Call Processing’spaceChng PL 0Chng PL 1Chng PL 2Chng PL 3Chng PL 4Chng PL 5Chng PL 6Chng PL 7MaintenAlertReturns quoted string?’:ORDer(Order):CPR

Strona 30

125Call Processing(Chan:):CPRocess:CALLP:RCDD1:RCDDATA1Returns quoted string?:RCDD2:RCDDATA2Returns quoted string?:RCDD3:RCDDATA3Returns quoted strin

Strona 31

126 Call Processing:CPRocess:CALLP:RECB(Word B):F:FWORDReturns quoted string?:NAWComingReturns quoted string?:LOCalReturns quoted string?:ORDQualifie

Strona 32 - Figure 1 Manual Control Mode

127Call Processing:CPRocess:CALLP:RECD(Word D):F:FWORDReturns quoted string?:NAWComingReturns quoted string?:DIG1:DIGIT1Returns quoted string?:DIG2:D

Strona 33

128 Call Processing:CPRocess:CALLP:RECE(Word E):F:FWORDReturns quoted string?:NAWComingReturns quoted string?:DIG9:DIGIT9Returns quoted string?:DIG10

Strona 34 - Getting Started

129Call Processing:CPRocess:CALLP:RCOConfirm(Order Confirmation Message):F:FWORDReturns quoted string?:NAWComingReturns quoted string?:LOCalReturns q

Strona 35

Contents 131 Using GPIBOverview of the Test Set 26Getting Started 34Remote Operation 47Addressing 49IEEE 488.1 Remote Interface Message Capabiliti

Strona 36

130 Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:SPOM1:SPOMESSAGE1(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredRe

Strona 37

131Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:SPOM2:SPOMESSAGE2(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredRet

Strona 38

132 Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:ACCess(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredReturns quote

Strona 39

133Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:RINCrement(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredReturns qu

Strona 40

134 Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:RIDentify(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredReturns qu

Strona 41

135Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:CFMessage(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredReturns quo

Strona 42 - Control Annunciators

136 Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:MSWord(T1T2):TYPE:T1T2:DCCode(DCC)’space2 chars requiredReturns quote

Strona 43

137Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:MSORder(T1T2):TYPE:T1T2:SCCode(SCC)’space2 chars requiredReturns quote

Strona 44

138 Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:MSVoice(T1T2):TYPE:T1T2:SCCode(SCC)’space2 chars requiredReturns quot

Strona 45 - Basic Programming Examples

139Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:FVORder(T1T2):TYPE:T1T2:SCCode(SCC)’space2 chars requiredReturns quote

Strona 46

Contents14 2 Methods For Reading Measurement Results Background 58HP® BASIC ‘ON TIMEOUT’ Example Program 60HP® BASIC ‘MAV’ Example Program 64

Strona 47 - Remote Operation

140 Call Processing’space2 chars requiredReturns quoted string?’:CPRocess:CALLP:FVVoice(T1T2):TYPE:T1T2:SCCode(SCC)’space2 chars requiredReturns quot

Strona 48 - Remote Capabilities

141DecoderDecoder For Decoder measurements see the MEASure command diagram.For selecting Decoder Input, see AF Anaylzer command diagram.See Real Numb

Strona 49 - Addressing

142S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\dec.sec Decoder:NAMPs or :NTACs and :EDACs’spaceNAMPSNTACSReturns quoted string?’:DECoder:STANDard’sp

Strona 50

143Decoder:AMPs or :TACs and :CDCSs, :DTMF, :FGEN, and :DPAG:DECoder:AMPS:TACS:PATTern:TRIGger’spacestringReturns quoted string?’:GATESee Real Number

Strona 51

144S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\dec.sec Decoder:TSEQuential, :MPT1327, and :LTR:DECoder:TSEQuential:GATESee Real Number Setting Synta

Strona 52

145DisplayDisplay:DISPlay:DISP continuedspaceACNTrolACPowerAFANalyzerCANanlyzerCBITCCNFigureCDANalyzerCDPDCDATaCDMAtestCGENeratorCMEasureCONFigureDEC

Strona 53 - Remote/Local Modes

146S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\disp.sec Display:DISPlayspaceTDMA testTESTsTFReqTHLPTIBasicTMAKeTPARmTPRintTSEQnTSPecReturns current

Strona 54 - Remote To Local Transitions

147MeasureMeasure :MEASure:MEAS continued:LRATio*Does not include the :STATe command’spaceSINADDistnSNRAF FreqDC LevelCurrentReturns quoted string?’:

Strona 55 - Clear Lockout/Set Local

148S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\meas.sec Measure:MEASure:MEAS continued:MARKer*Does not include the :METer command:TRACe?Returns real

Strona 56

149Measure:MEASure:MEAS:DEC Continued On Next Page:DECoder?Returns integer value?Returns real valueSee Number Measurement Syntax*:CDCSs:RATE:BITS:COD

Strona 57 - Results

Contents 153 GPIB Command Guidelines Sequential and Overlapped Commands 70Guidelines for Operation 71

Strona 58 - Background

150S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\meas.sec Measure:MEASure:DECoder:DTMF:LOW:HIGH:FREQuency:ABSolute:ERRor?Returns up to 19 real valuesS

Strona 59

151Measure:MEASure:DECoder:NAMPs:LOW:HIGH:FREQuency:ABSolute:ERRor?Returns up to 17 real valuesSee Multiple Number Measurement Syntax?Returns up to 1

Strona 60

152S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\meas.sec Measure:MEASure:DECoder:MEAS:DEC Continued On Next Page:MPT1327?Returns quoted string:BUFFer

Strona 61

153Measure:MEASure:DECoder:LTR:RATE:DATA?Returns real value:EDACs?Returns quoted string of 74 characters?Returns quoted stringSee Number Measurement

Strona 62

154S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\osc.sec OscilloscopeOscilloscope For Oscilloscope measurements see the MEASure command diagram.:OSCil

Strona 63

155Oscilloscope:OSCilloscope’space200 ms100 ms50 ms20 ms10 ms5 ms2 ms1 ms500 uS200 uS100 uS50 uS20 uS10 uS5 uS2 uS1 uSReturns quoted string?’:SCALe:T

Strona 64 - BASIC ‘MAV’ Example Program

156S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\osc.sec Oscilloscope:OSCilloscope’space50 %20 %10 %5 %2%1%0.5 %0.2%0.1%0.05 %Returns quoted string?’:

Strona 65

157Oscilloscope:OSCilloscope’space20 V10 V5 V2 V1 V500 mV200 mV100 mV50 mV20 mV10 mV5 mV2 mV1 mV500 uV200 uV100 uV50 uV20 uVReturns quoted string?’:S

Strona 66

158S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\osc.sec Oscilloscope:OSCilloscope:LEVel:TRIGger:PRETriggerSee Real Number Setting Syntax**Does not in

Strona 67

159ProgramProgram :PROGram:SELected:ALL:DELete:DEFinespaceprogramReturns <length info><program>?:EXECutestringspaceCONTinuePAUSeRUNSTOPRe

Strona 68

Contents16 4 GPIB Commands GPIB Syntax Diagrams 92Adjacent Channel Power (ACP) 95AF Analyzer 97AF Generator 1 100AF Generator 2 Pre-Modulation Fil

Strona 69 - Command Guidelines

160S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\reg.sec Save/Recall RegistersSave/Recall Registers NOTE:1 The Test Set does not check for a duplicat

Strona 70

161RF AnalyzerRF Analyzer For RF Analyzer measurements see the MEASure command diagram.:RFANalyzer:RFAN continued:ZERO:PMEasurement’space40 dB20 dB0

Strona 71 - Guidelines for Operation

162S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\rfa.sec RF Analyzer:RFANalyzer(Tune Mode)’spaceNormalHighReturns quoted string?’:SENSitivity’spaceOnO

Strona 72 - Command Punctuation

163RF GeneratorRF Generator For RF Generator measurements see the MEASure command diagram.1 RF power must not be applied while zeroing. Set the RF GE

Strona 73 - AFAN:INP:GAIN ’40 dB’

164S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\rin.sec Radio InterfaceRadio Interface :RINTerface:STATus:INTerrupt2:SEND’spaceArmDisableReturns quot

Strona 74 - ENTER 714;Afg1_to$

165Spectrum AnalyzerSpectrum Analyzer For Spectrum Analyzer measurements see the MEASure command diagram.:SANalyzer:SAN continued:CFREquency:MARKer’s

Strona 75

166S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\san.sec Spectrum Analyzer:SANalyzer’spaceTrackFixedReturns quoted string?’:RFGenerator:RLEVelSee Real

Strona 76 - MEAS:RFR:POW:DUN DBM

167GPIB Only CommandsGPIB Only Commands:SPECial:TOTALUSERRAMReturns integer string?:RAMFORIBASICReturns integer string?:RAMDISKALLOCReturns integer s

Strona 77 - ENTER 714;A$

168S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\stat.sec StatusStatus:STATus:PRESet:EVENt?Returns integer value:CONDition:ENABle:NTRansition:PTRansit

Strona 78

169SystemSystem:SYSTem:ERRorReturns integer value, quoted string?

Strona 79

Contents 17Status 168System 169Tests 170Trigger 173Integer Number Setting Syntax 174Real Number Setting Syntax 175Multiple Real Number Setting Synt

Strona 80

170S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\test.sec TestsTests :TESTs:CONFigure,spaceinteger valueReturns unquoted string, 5 elements seperated

Strona 81

171TestsReturns unquoted string, 2 elements seperated by commas:TESTs:FREQuencyspaceReturns unquoted string, 7 elements seperated by commas?integer v

Strona 82

172S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\test.sec Tests:TESTs’UpperLowerBothNone’:NUMBer:SEQNumberReturns unquoted string, 3 elements seperate

Strona 83

173TriggerTrigger :TRIGger:RETRigger:MODE:IMMediate’spaceFASTFULLReturns FAST or FULL?’:SETTling’spaceREPetitiveSINGleReturns REP or SING?’:ABORt

Strona 84

174S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\inss.sec Integer Number Setting SyntaxInteger Number Setting Syntax :INCRementPrevious Syntax?Returns

Strona 85

175Real Number Setting SyntaxReal Number Setting Syntax :INCRement:DUNits:MODE?Returns LIN or LOGspace LINearLOGarithmPrevious Syntax?Returns real va

Strona 86

176S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\mrnss.sec Multiple Real Number Setting SyntaxMultiple Real Number Setting Syntax :INCRement:DUNits:MO

Strona 87 - Using the STATe Command

177Number Measurement SyntaxNumber Measurement Syntax :DUNits?Returns GPIB unitsspaceGPIB units?Returns unitsspaceunits:STATe?Returns 1 or 0space 1ON

Strona 88 - ENTER 714;State_on_off

178S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\nms.sec Number Measurement SyntaxPrevious Syntax:UNITs?Returns GPIB unitsspaceGPIB units?Returns 1 or

Strona 89 - Sample GPIB Program

179Multiple Number Measurement SyntaxMultiple Number Measurement Syntax :DUNitsPrevious Syntax?Returns unitsspaceunits:UNITs?Returns GPIB unitsspaceG

Strona 90

Contents18 5 Advanced Operations Increasing Measurement Throughput 234Status Reporting 239GPIB Service Requests 293Instrument Initialization 303

Strona 91 - GPIB Commands

180S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsEquivalent Front-Panel Key Commands Most front-panel key

Strona 92 - GPIB Syntax Diagrams

181Equivalent Front-Panel Key CommandsDATA FUNCTIONS Keys The Data Functions keys can be divided into two groups; those which affect measurements (RE

Strona 93

182S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsGuidelines for Using Numeric Entry Field Data Functions

Strona 94 - Diagram Conventions

183Equivalent Front-Panel Key CommandsTo Turn Measurement Averaging ON and OFF. Use the :AVERage:STATe commands to turn the averaging data function

Strona 95 - Adjacent Channel Power (ACP)

184S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Reset Averaging. Use the :AVERage:RESet commands to

Strona 96

185Equivalent Front-Panel Key CommandsHI LIMIT and LO LIMIT The HI LIMIT and LO LIMIT Data Functions are used to define a measurement “window” which

Strona 97 - AF Analyzer

186S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Query the State of High and Low Measurement Limit Che

Strona 98

187Equivalent Front-Panel Key CommandsTo Set the Display Units for High and Low Measurement Limits. Use the :HLIMit:DUNits and :LLIMit:DUNits comma

Strona 99

188S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Query the High and Low Measurement Limit Settings.

Strona 100 - AF Generator 1

189Equivalent Front-Panel Key CommandsTo Reset Measurement Limit Detection. Use the :HLIMit:RESet and :LLIMit:RESet commands to reset measurement l

Strona 101

Contents 196 Memory Cards/Mass Storage Default File System 324Mass Storage Device Overview 325Default Mass Storage Locations 331Mass Storage Acce

Strona 102 - AF Generator 2/Encoder

190S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Query the Increment Value. Use the :INCRement? comm

Strona 103 - :AMPS or :TACS

191Equivalent Front-Panel Key CommandsTo Set the Increment Value Display Units. Use the :INCRement:DUNits commands to set the units-of-measure used

Strona 104

192S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsINCR×10 The INCR×10 Data Function increases the incremen

Strona 105 - :NAMPs or :NTACs

193Equivalent Front-Panel Key CommandsIncrement Up/Down (Arrow Keys) The Increment Up/Down (Arrow Keys) Data Functions change the field’s setting by

Strona 106

194S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Query the State of the Meter. Use the :METer:STATe?

Strona 107 - :CDCSs and :DTMF

195Equivalent Front-Panel Key CommandsTo Set the Meter High End and Low End Points. Use the :METer:HEND and :MEter:LEND commands to set the analog

Strona 108 - :DPAGing

196S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Set the Meter High End and Low End Point Display Unit

Strona 109

197Equivalent Front-Panel Key CommandsREF SET The REF SET Data Function establishes a measurement reference point. The GPIB command :REFerence is use

Strona 110 - :FGENerator and :TSEQuential

198S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTo Set A Measurement Reference Point. Use the :REFerenc

Strona 111

199Equivalent Front-Panel Key CommandsTo Set Measurement Reference Point Display Units. Use the :REFerence:DUNits commands to set a measurement ref

Strona 112

2 © Copyright Agilent Technologies 1997, 2000Notice No part of this manual may be reproduced in any form or by any means (including electronic storag

Strona 113

Contents20 7 IBASIC Controller Introduction 354The IBASIC Controller Screen 355Important Notes for Program Development 357Program Development 358

Strona 114

200S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsINSTRUMENT STATE Keys ADRS The ADRS key displays the Tes

Strona 115 - :MPT1327

201Equivalent Front-Panel Key CommandsLOCAL The LOCAL key returns the Test Set to local, front-panel control. The Test Set returns to Local operatio

Strona 116 - 2 String maximum length 300

202S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsRECALL The RECALL key is used to recall an instrument s

Strona 117 - Configure, I/O Configure

203Equivalent Front-Panel Key CommandsRemoving Saved Instrument States. One or all of the saved instrument states can be removed from the selected

Strona 118

204S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsSCREEN CONTROL Keys and To Screen FieldIn manual mode, t

Strona 119 - (RFGen Volts)

205Equivalent Front-Panel Key CommandsHOLD The HOLD key is used to hold/resume all active measurements. There is no equivalent GPIB command for the

Strona 120

206S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\efpkc.sec Equivalent Front-Panel Key CommandsTable 13 Screen Mnemonics for the DISPlay CommandMnemoni

Strona 121

207Equivalent Front-Panel Key Commands

Strona 122 - Call Processing

208S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common CommandsIEEE 488.2 Common Commands The IEEE 488.2 Standard defines a set o

Strona 123

209IEEE 488.2 Common CommandsCommon Command Descriptions *IDN? (Identification Query) The *IDN? query causes a device to send its identification info

Strona 124

Contents 218 Programming the Call Processing SubsystemDescription of the Call Processing Subsystem’s Remote User Interface 426Using the Call Process

Strona 125

210S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common Commands*OPT? (Option Identification Query) The *OPT? command tells the Te

Strona 126

211IEEE 488.2 Common Commands*RST (Reset) The *RST command resets the Test Set. When the *RST command is received the majority of fields in the Test

Strona 127

212S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common Commands*TST? (Self-Test Query) The *TST? self-test query causes the Test

Strona 128

213IEEE 488.2 Common Commands*OPC (Operation Complete) The *OPC command allows for synchronization between the Test Set and an external controller. T

Strona 129

214S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common CommandsCAUTION: The *OPC? command should not be used for determining if a

Strona 130

215IEEE 488.2 Common CommandsExample program: Using *OPC through polling of the Standard Event Status Register 10 INTEGER Stdevnt_reg_val20 OUTPUT 71

Strona 131

216S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common Commands*OPC? (Operation Complete Query) The *OPC? query allows for synchr

Strona 132

217IEEE 488.2 Common CommandsUsing the *OPC? query by reading Output Queue Bit 4 in the Service Request Enable Register is set to a value of zero (di

Strona 133

218S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common CommandsExample program 10 OUTPUT 714;"*SRE 16" !

Strona 134

219IEEE 488.2 Common Commands*WAI (Wait To Complete) The *WAI command stops the Test Set from executing any further commands or queries until all com

Strona 135

Contents22 9 Error Messages

Strona 136

220S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common Commands*CLS (Clear Status) The *CLS command clears the contents (sets all

Strona 137

221IEEE 488.2 Common Commands*PCB (Pass Control Back) The Test Set accepts the *PCB command. Refer to “Passing Instrument Control” in the Advanced Op

Strona 138

222S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\ieee.sec IEEE 488.2 Common Commands*RCL (Recall Instrument State) The *RCL command restores the state

Strona 139

223IEEE 488.2 Common Commands

Strona 140

224S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\tm.sec Triggering MeasurementsTriggering Measurements The measurement cycle is started (triggered) by

Strona 141 - Decoder

225Triggering MeasurementsTrigger Modes The Trigger Mode is defined by two parameters: retriggering and settling. Retriggering Retriggering refers to

Strona 142 - :NAMPs or :NTACs and :EDACs

226S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\tm.sec Triggering MeasurementsSettling Settling refers to the amount of delay introduced to allow sig

Strona 143

227Triggering MeasurementsDefault Trigger Mode The Trigger mode is set to FULL SETTling and REPetitive RETRiggering whenever• the Test Set is powered

Strona 144

228S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\tm.sec Triggering MeasurementsTrigger Commands :TRIGger:IMMediate The :TRIGger:IMMediate command tell

Strona 145

229Triggering Measurements:MODE The :MODE command is used to set the Trigger Mode for all active measurements. The trigger mode is defined by two par

Strona 146

Contents 23Index 569

Strona 147 - Measure

230S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\tm.sec Triggering MeasurementsTrigger Mode and Measurement Speed There are two generalized scenarios

Strona 148

231Triggering MeasurementsMeasurement Pacing Measurement pacing can be accomplished by using the IEEE 488.2-1987 Common Commands *OPC, *OPC?, and *WA

Strona 149

232S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\tm.sec Triggering Measurements

Strona 150

2335 Advanced Operations

Strona 151

234S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsIncreasing Measurement ThroughputIncreasing Measurement Thro

Strona 152

235Chapter 5, Advanced OperationsIncreasing Measurement ThroughputTime-to-first-reading after making new settings is usually much slower than the rep

Strona 153

236S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsIncreasing Measurement Throughput3. Frequency Counter Gate T

Strona 154 - Oscilloscope

237Chapter 5, Advanced OperationsIncreasing Measurement ThroughputOptimizing the Execution Speed of the Control Program Execution speed of the contro

Strona 155

238S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsIncreasing Measurement ThroughputExample #3OUTPUT 714;"

Strona 156

239Chapter 5, Advanced OperationsStatus ReportingStatus Reporting This section describes the status reporting structure used in the Test Set. The str

Strona 157

Contents24

Strona 158

240S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingFigure 3 Status Reporting Structure Overvie

Strona 159 - Program

241Chapter 5, Advanced OperationsStatus ReportingStatus Byte Register The Status Byte Register is an 8-bit register that is used to summarize the Sum

Strona 160 - Save/Recall Registers

242S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingFigure 4 Status Byte RegisterTable 17 detai

Strona 161 - RF Analyzer

243Chapter 5, Advanced OperationsStatus ReportingThe Status Byte Register is unique in that bit 6 can take on two different meanings. The contents of

Strona 162

244S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingReading with a Serial Poll The contents of t

Strona 163 - RF Generator

245Chapter 5, Advanced OperationsStatus ReportingExample BASIC program to read Status Byte with *STB command10 INTEGER Stat_byte_reg,Stat_byte,Mstr_s

Strona 164 - Radio Interface

246S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingFigure 5 Status Data Structure - Register M

Strona 165 - Spectrum Analyzer

247Chapter 5, Advanced OperationsStatus ReportingStatus Register Model This section explains how the status registers are structured in the Test Set.

Strona 166

248S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingTransition Filters are read-write. Transitio

Strona 167 - GPIB Only Commands

249Chapter 5, Advanced OperationsStatus ReportingStatus Reporting Structure Operation. In general the status reporting structure described on the p

Strona 168

251 Using GPIB11. GPIB was formerly called HP-IB for Hewlett-Packard instruments. Some labels on the instrument may still reflect the former HP® name

Strona 169

250S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingStatus Queue Model This section explains how

Strona 170

251Chapter 5, Advanced OperationsStatus ReportingStatus Register Group Contents Figure 7 shows the Status Register Groups in the Test Set. The conten

Strona 171

252S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingOperation Status Register Group The Operatio

Strona 172

253Chapter 5, Advanced OperationsStatus ReportingTable 18 shows the Operation Status Register Group Condition Register bit assignments. Table 18 Oper

Strona 173 - Trigger

254S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingAccessing the Operation Status Register Grou

Strona 174 - Integer Number Setting Syntax

255Chapter 5, Advanced OperationsStatus ReportingReading the Event Register SyntaxSTATus:OPERation:EVENt?ExampleOUTPUT 714;"STAT:OPER:EVEN?&

Strona 175 - Real Number Setting Syntax

256S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingStandard Event Status Register Group The Sta

Strona 176

257Chapter 5, Advanced OperationsStatus ReportingAccessing the Standard Event Status Register Group’s Registers Table 19 Standard Event Status Regist

Strona 177 - Number Measurement Syntax

258S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingThe following sections show the syntax and g

Strona 178

259Chapter 5, Advanced OperationsStatus ReportingExampleOUTPUT 714;"*ESR?"ENTER 714;Register_valueThe *ESR? query allows the programmer to

Strona 179

26S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBOverview of the Test SetOverview of the Test Set The Test Set combines

Strona 180 - DATA Keys

260S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingReading the Enable Register Syntax*ESE?Ex

Strona 181 - DATA FUNCTIONS Keys

261Chapter 5, Advanced OperationsStatus ReportingWriting the Enable Register Syntax*ESE <integer>ExampleOUTPUT 714;"*ESE 255"The *E

Strona 182

262S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingOutput Queue Group The Output Queue Group is

Strona 183 - Use the :AVERage:STAT?

263Chapter 5, Advanced OperationsStatus ReportingAccessing the Output Queue When messages are sent to the Test Set, it decodes the message to determi

Strona 184

264S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingError Message Queue Group The Error Message

Strona 185

265Chapter 5, Advanced OperationsStatus ReportingAccessing the Error Message Queue A message appears in the Error Message Queue any time bit 2, 3, 4,

Strona 186

266S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingQuestionable Data/Signal Register Group The

Strona 187

267Chapter 5, Advanced OperationsStatus ReportingFigure 12 Questionable Data/Signal Register Group0:STATus: QUEStionable: ENABle?<integer>:STA

Strona 188

268S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingTable 20 shows the Questionable Data/Signal

Strona 189

269Chapter 5, Advanced OperationsStatus ReportingAccessing the Questionable Data/Signal Register Group’s Registers The following sections show the sy

Strona 190

27Chapter 1, Using GPIBOverview of the Test SetManual Control Mode The Test Set’s primary instruments are shown on the left side of Figure 1. There a

Strona 191 - Generator’s frequency

270S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingReading the Event Register SyntaxSTATus:Q

Strona 192 - :INCRement:DIVide

271Chapter 5, Advanced OperationsStatus ReportingCall Processing Status Register Group The Call Processing Status Register Group contains information

Strona 193 - :INCRement <UP or DOWN>

272S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingFigure 13 Call Processing Status Register G

Strona 194

273Chapter 5, Advanced OperationsStatus ReportingTable 21 Call Processing Status Register Group, Condition Register Bit Assignments Bit Number Binar

Strona 195

274S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingAccessing the Call Processing Status Registe

Strona 196

275Chapter 5, Advanced OperationsStatus ReportingReading the Event Register SyntaxSTATus:CALLProc:EVENt?ExampleOUTPUT 714;"STAT:CALLP:EVEN?&

Strona 197 - Power measurement

276S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingCalibration Status Register Group The Calibr

Strona 198 - ENTER 714;Ref_val

277Chapter 5, Advanced OperationsStatus ReportingTable 22 details the Calibration Status Register Group’s Condition Register bit assignments. Table 2

Strona 199 - TX Power measurement

278S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingAccessing the Calibration Status Register Gr

Strona 200 - INSTRUMENT STATE Keys

279Chapter 5, Advanced OperationsStatus ReportingClearing the Event Register The EVENT register is cleared whenever it is queried or whenever the

Strona 201 - OUTPUT 714;"*RST"

28S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBOverview of the Test SetInternal Automatic Control Mode In Internal Au

Strona 202

280S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingHardware Status Register #2 Group The Hardwa

Strona 203 - RECALL mass storage device

281Chapter 5, Advanced OperationsStatus ReportingTable 23 shows the Hardware Status Register Group #2’s Condition Register bit assignments. Table 23

Strona 204 - ENTER 714;Disp_screen$

282S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingAccessing the Hardware Status Register #2 Gr

Strona 205 - USER Keys

283Chapter 5, Advanced OperationsStatus ReportingReading the Event Register SyntaxSTATus:HARDware2:EVENt?ExampleOUTPUT 714;"STAT:HARD2:EVEN?

Strona 206

284S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingHardware Status Register #1 Group The Hardwa

Strona 207

285Chapter 5, Advanced OperationsStatus ReportingTable 24 shows the Hardware Status Register Group #1’s Condition Register bit assignments. Table 24

Strona 208 - IEEE 488.2 Common Commands

286S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingAccessing the Hardware Status Register #1 Gr

Strona 209 - (Identification

287Chapter 5, Advanced OperationsStatus ReportingReading the Transition Filters SyntaxSTATus:HARDware1:PTRansition?STATus:HARDware1:NTRansition?E

Strona 210 - Identification

288S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingClearing the Event Register The EVENT reg

Strona 211

289Chapter 5, Advanced OperationsStatus ReportingCommunicate Status Register Group The Communicate Status Register Group contains information about t

Strona 212 - *TST? (Self-Test

29Chapter 1, Using GPIBOverview of the Test SetDifferences Between the Test Set’s IBASIC Controller and Other Single-Tasking ControllersThe IBASIC Co

Strona 213 - Complete)

290S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingTable 25 shows the Communicate Status Regist

Strona 214

291Chapter 5, Advanced OperationsStatus ReportingAccessing the Communicate Status Register Group’s Registers The following sections show the syntax a

Strona 215

292S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsStatus ReportingReading the Event Register SyntaxSTATus:C

Strona 216 - Complete Query)

293Chapter 5, Advanced OperationsGPIB Service RequestsGPIB Service Requests The Test Set is capable of generating a “service request” when it require

Strona 217

294S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsGPIB Service RequestsSetting Up and Enabling SRQ Interrupts

Strona 218 - Example program

295Chapter 5, Advanced OperationsGPIB Service RequestsService Request Enable Register Service request enabling allows the application programmer to s

Strona 219 - *WAI (Wait To

296S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsGPIB Service RequestsReading the Service Request Enable Regi

Strona 220

297Chapter 5, Advanced OperationsGPIB Service RequestsExampleOUTPUT 714;"*SRE 18"NOTE: The decimal value of the bit pattern must be a posit

Strona 221

298S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsGPIB Service RequestsProcedure for Generating a Service Requ

Strona 222 - SAVE/RECALL field on

299Chapter 5, Advanced OperationsGPIB Service RequestsExample Program to Set Up and Service an SRQ Interrupt The following Instrument BASIC program w

Strona 223

3Safety Summary The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with

Strona 224 - Triggering Measurements

30S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBOverview of the Test SetExternal Automatic Control Mode In External Au

Strona 225 - Trigger Modes

300S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsGPIB Service Requests410 ! Bit 0: Spec Anal Self-cal fai

Strona 226

301Chapter 5, Advanced OperationsGPIB Service Requests900 ! decimal value = 2^5 = 32910 ! Bit 3 = Q

Strona 227 - Default Trigger Mode

302S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsGPIB Service Requests1400 !1410 ENABLE INTR 7;21420SUBEXIT1

Strona 228 - Trigger Commands

303Chapter 5, Advanced OperationsInstrument InitializationInstrument Initialization This section discusses the various methods available to the progr

Strona 229

304S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsInstrument InitializationMethods of Initialization There are

Strona 230

305Chapter 5, Advanced OperationsInstrument InitializationTable 28 Screen Fields Restored/Initialized During Power-On Reset CONFIGURE Screen Fields

Strona 231 - Measurement Pacing

306S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsInstrument InitializationThe Power-On Reset condition in the

Strona 232

307Chapter 5, Advanced OperationsInstrument InitializationTable 29 Screen Fields Restored/Initialized During Front Panel Reset CONFIGURE Screen Field

Strona 233 - Advanced Operations

308S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsInstrument InitializationThe Front-panel Reset condition in

Strona 234 - Optimizing Measurement Speed

309Chapter 5, Advanced OperationsInstrument InitializationTable 30 Screen Fields Restored/Initialized During *RST Reset CONFIGURE Screen Fields PRIN

Strona 235 - :AFAN:RANG ’HOLD’

31Chapter 1, Using GPIBOverview of the Test SetWriting programs for the Test Set One of the design goals for automatic control of the Test Set was th

Strona 236 - :RFAN:GTIM <value> MS

310S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsInstrument InitializationThe *RST Reset condition in the Tes

Strona 237 - ON"

311Chapter 5, Advanced OperationsInstrument InitializationDevice Clear (DCL) GPIB Bus Command The Device Clear (DCL) Reset is accomplished by sending

Strona 238 - ENTER 714;Power,Freq_abs

312S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsInstrument InitializationSelected Device Clear (SDC) GPIB Bu

Strona 239 - Status Reporting

313Chapter 5, Advanced OperationsPassing ControlPassing Control Communications on the GPIB bus are accomplished according to a precisely defined set

Strona 240 - Structure

314S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsPassing ControlConfiguring the Test Set as the System Contro

Strona 241 - Status Byte Register

315Chapter 5, Advanced OperationsPassing ControlPassing Control to the Test Set Control is passed to the Test Set when it is addressed to TALK and th

Strona 242 - Read by STB?

316S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsPassing ControlPassing Control Back to Another Controller Th

Strona 243

317Chapter 5, Advanced OperationsPassing ControlRequesting Control using IBASIC The Test Set has the capability to request control of the bus from th

Strona 244

318S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsPassing Control1. If a program is running on the HP® 9000 Wo

Strona 245

319Chapter 5, Advanced OperationsPassing ControlThe following program would run in the External Controller: 10 COM /Gpib_names/ INTEGER Internal_g

Strona 246

32S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBOverview of the Test SetFigure 1 Manual Control ModeCRTKEYPAD/ROTARY

Strona 247

320S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsPassing Control480 ! Data is ready in the Test Set490 OU

Strona 248

321Chapter 5, Advanced OperationsPassing ControlThe following IBASIC program would be loaded off the Memory Card and run in the Test Set: 10 COM

Strona 249

322S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\advoper.fb Chapter 5, Advanced OperationsPassing Control600 OUTPUT Internal_gpib;"DISP TX;M

Strona 250

3236 Memory Cards/Mass Storage This chapter contains information about using the mass storage devices available in the Test Set for storing and retri

Strona 251

324S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageDefault File SystemDefault File System Default File Sy

Strona 252

325Chapter 6, Memory Cards/Mass StorageMass Storage Device OverviewMass Storage Device Overview As shown in Figure 19 on page 326, the Test Set has b

Strona 253

326S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageMass Storage Device OverviewFigure 19 Internal and Ex

Strona 254

327Chapter 6, Memory Cards/Mass StorageMass Storage Device OverviewPrograms and data can be retrieved from any of these mass storage devices. Program

Strona 255

328S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageMass Storage Device Overview Typical Uses• Permanent s

Strona 256

329Chapter 6, Memory Cards/Mass StorageMass Storage Device OverviewTypical Uses • Semi-permanent program and data storage • Semi-permanent Save/Recal

Strona 257

33Chapter 1, Using GPIBOverview of the Test SetFigure 2 Internal Automatic Control ModeCRTKEYPAD/ROTARY KNOBTO SCREENCONTROLHARDWAREINSTRUMENT CON-T

Strona 258

330S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageMass Storage Device OverviewTypical Uses • Permanent s

Strona 259

331Chapter 6, Memory Cards/Mass StorageDefault Mass Storage LocationsDefault Mass Storage Locations Built-in IBASIC Controller The default mass stora

Strona 260

332S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageDefault Mass Storage LocationsExternal Disk Drive The

Strona 261 - *ESE <integer>

333Chapter 6, Memory Cards/Mass StorageMass Storage AccessMass Storage Access Program and data files stored on the Test Set’s various mass storage lo

Strona 262 - Output Queue Group

334S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageDOS and LIF File System ConsiderationsDOS and LIF File

Strona 263 - Enter 714;Output_data

335Chapter 6, Memory Cards/Mass StorageDOS and LIF File System ConsiderationsThe period (.) may appear in the name but only to separate the file name

Strona 264 - Error Message Queue Group

336S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageDOS and LIF File System ConsiderationsPotential File N

Strona 265

337Chapter 6, Memory Cards/Mass StorageDOS and LIF File System Considerations• When copying LIF named files to a DOS formatted media, the file name i

Strona 266

338S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageDOS and LIF File System ConsiderationsInitializing Med

Strona 267

339Chapter 6, Memory Cards/Mass StorageDOS and LIF File System ConsiderationsTESTS Subsystem DOS File Restrictions The Test Set uses IBASIC revision

Strona 268

34S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting StartedGetting Started What is GPIB? The General Purpose Inter

Strona 269

340S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageUsing the ROM DiskUsing the ROM Disk The Test Set come

Strona 270

341Chapter 6, Memory Cards/Mass StorageUsing Memory CardsUsing Memory Cards OTP (One Time Programmable) cards provide removable read-only storage. Fi

Strona 271

342S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageUsing Memory CardsNOTE: Memory cards may be inserted a

Strona 272

343Chapter 6, Memory Cards/Mass StorageUsing Memory CardsSetting the Write-Protect Switch The SRAM memory card’s write-protect switch lets the user s

Strona 273

344S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageUsing Memory CardsThe Memory Card Battery SRAM memory

Strona 274

345Chapter 6, Memory Cards/Mass StorageUsing Memory CardsMemory Card Mass Storage Volume Specifier The front-panel memory card slot’s mass storage vo

Strona 275

346S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageBacking Up Procedure and Library FilesBacking Up Proce

Strona 276

347Chapter 6, Memory Cards/Mass StorageCopying Files Using IBASIC CommandsCopying Files Using IBASIC Commands Files can be copied from one mass stora

Strona 277

348S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageCopying Files Using IBASIC CommandsNOTE: Using the vol

Strona 278

349Chapter 6, Memory Cards/Mass StorageUsing RAM DiskUsing RAM Disk RAM Disk is a section of the Test Set’s internal RAM memory that has been set asi

Strona 279 - ENTER 714;Register_value

35Chapter 1, Using GPIBGetting StartedGPIB Information Provided in This Manual What Is Explained • How to configure the Test Set for GPIB operation •

Strona 280

350S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageUsing RAM DiskInitializing RAM Disks Each RAM Disk uni

Strona 281 - Too Wide

351Chapter 6, Memory Cards/Mass StorageUsing External Disk DrivesUsing External Disk Drives The Test Set supports only GPIB external disk drives. Cer

Strona 282

352S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\memcard.fb Chapter 6, Memory Cards/Mass StorageUsing External Disk Drives

Strona 283

3537 IBASIC Controller

Strona 284

354S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerIntroductionIntroduction The Test Set contains an instrument co

Strona 285

355Chapter 7, IBASIC ControllerThe IBASIC Controller ScreenThe IBASIC Controller Screen The Test Set has a dedicated screen for interfacing with the

Strona 286 - STATus:HARDware1:CONDition?

356S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerThe IBASIC Controller ScreenThe TESTS (IBASIC Controller) scree

Strona 287

357Chapter 7, IBASIC ControllerImportant Notes for Program DevelopmentImportant Notes for Program Development The Test Set is designed to operate the

Strona 288 - STATus:HARDware1:ENABle?

358S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerProgram DevelopmentProgram Development There are three recommen

Strona 289

359Chapter 7, IBASIC ControllerProgram DevelopmentFigure 24 Program Development MethodsDevelop on Test Setusing screen"EDIT" modeRun progr

Strona 290

36S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting StartedGeneral GPIB Programming Guidelines The following guide

Strona 291

360S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsInterfac

Strona 292

361Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsSerial Port 9. Serial Port 9 is used for developing and editi

Strona 293 - GPIB Service Requests

362S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsTable 39

Strona 294

363Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsFigure 25 Available Agilent RS-232 Serial Cables and Adapters0

Strona 295 - , is an 8-bit register that

364S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsFigure 2

Strona 296

365Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsSerial Port 9 Configuration Table 41 on page 366 and the follow

Strona 297

366S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsReceive

Strona 298

367Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsPC Configuration To prepare for IBASIC program development, the

Strona 299

368S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsSetting

Strona 300

369Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsExample Terminal Communications Settings• Baud Rate: 9600• Data

Strona 301

37Chapter 1, Using GPIBGetting Started2. When entering the TX TEST screen, a. The AF Anl In field, the De-Emphasis field, the Detector field and the

Strona 302

370S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial Ports5. Press

Strona 303 - Instrument Initialization

371Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial PortsSetting Up Agilent AdvanceLink (Agilent 68333F Version B.02.00)

Strona 304 - Power-On Reset

372S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerInterfacing to the IBASIC Controller using Serial Ports6. Press

Strona 305

373Chapter 7, IBASIC ControllerChoosing Your Development MethodChoosing Your Development Method There are three fundamental methods for developing IB

Strona 306 - Front-panel PRESET Key

374S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerChoosing Your Development MethodMethod 1 Using a BASIC language

Strona 307

375Chapter 7, IBASIC ControllerMethod #1. Program Development on an External BASIC Language ComputerMethod #1. Program Development on an External BAS

Strona 308

376S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #1. Program Development on an External BASIC Language Co

Strona 309

377Chapter 7, IBASIC ControllerMethod #1. Program Development on an External BASIC Language ComputerHP® BASIC for Windows® PC Configuration for Windo

Strona 310

378S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #1. Program Development on an External BASIC Language Co

Strona 311

379Chapter 7, IBASIC ControllerMethod #1. Program Development on an External BASIC Language ComputerNOTE: The dedicated GPIB interface at select code

Strona 312

38S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting Started• Guideline #2. When developing programs to make measur

Strona 313 - Passing Control

380S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #1. Program Development on an External BASIC Language Co

Strona 314

381Chapter 7, IBASIC ControllerMethod #2. Developing Programs on the Test Set Using the IBASIC EDIT ModeMethod #2. Developing Programs on the Test Se

Strona 315

382S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #2. Developing Programs on the Test Set Using the IBASIC

Strona 316

383Chapter 7, IBASIC ControllerMethod #2. Developing Programs on the Test Set Using the IBASIC EDIT ModeEntering and Exiting the IBASIC EDIT Mode To

Strona 317 - Pass Control Examples

384S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #2. Developing Programs on the Test Set Using the IBASIC

Strona 318

385Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Least Preferred)Method #3. Developing Programs Using Word

Strona 319

386S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Le

Strona 320

387Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Least Preferred)Transferring Programs from the Word Proce

Strona 321

388S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Le

Strona 322

389Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Least Preferred)10 ! ASCII_DN20 ! Program to download A

Strona 323 - Memory Cards/Mass Storage

39Chapter 1, Using GPIBGetting Started7. Send the MEASure query command to initiate a reading. This will place the measured value into the Test Set’s

Strona 324 - Default File System

390S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Le

Strona 325 - Mass Storage Device Overview

391Chapter 7, IBASIC ControllerMethod #3. Developing Programs Using Word Processor on a PC (Least Preferred)Sending ASCII Text Files over RS-232 with

Strona 326

392S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerUploading Programs from the Test Set to a PCUploading Programs

Strona 327

393Chapter 7, IBASIC ControllerSerial I/O from IBASIC ProgramsSerial I/O from IBASIC Programs There are two serial ports available for I/O (input / o

Strona 328 - Typical Uses

394S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerSerial I/O from IBASIC ProgramsExample IBASIC Program Using Ser

Strona 329

395Chapter 7, IBASIC ControllerSerial I/O from IBASIC ProgramsSerial Port 10 Information Serial Port 10 is sometimes called Serial Port B in Test Set

Strona 330

396S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemPROGram Subsystem IntroductionThe PROGram Subs

Strona 331 - Built-in IBASIC Controller

397Chapter 7, IBASIC ControllerPROGram SubsystemSupported SCPI CommandsThe Test Set supports the following subset of the :SELected SCPI commands.• :S

Strona 332 - TESTS Subsystem

398S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemPROGram Subsystem Commands See the syntax diag

Strona 333 - Mass Storage Access

399Chapter 7, IBASIC ControllerPROGram SubsystemNOTE: It is possible for the PROGram Subsystem to think that there is an IBASIC program resident in t

Strona 334 - File Naming Conventions

4 GROUND THE INSTRUMENTTo minimize shock hazard, the instrument chassis and cover must be connected to an electrical protective earth ground. The ins

Strona 335

40S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting Started• Guideline #3. Avoid program hangs. If the program sto

Strona 336

400S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemSyntax (length of program known)PROGram[:SELec

Strona 337 - File Naming Recommendations

401Chapter 7, IBASIC ControllerPROGram SubsystemExample#16<data byte><data byte><data byte><data byte><data byte><da

Strona 338 - Storing Code Files

402S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram Subsystem:DELete:ALL The :DELete:ALL command is used t

Strona 339

403Chapter 7, IBASIC ControllerPROGram Subsystem:NUMber <varname>{,<nvalues>} The :NUMBer command is used to set, from an external cont

Strona 340 - Using the ROM Disk

404S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemExample setting the value of a two dimensional

Strona 341 - Using Memory Cards

405Chapter 7, IBASIC ControllerPROGram SubsystemSyntaxPROGram[:SELected]:NUMBer? <varname>NOTE: The program commands and syntax used to enter d

Strona 342

406S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemExample querying the value of a one dimensiona

Strona 343

407Chapter 7, IBASIC ControllerPROGram SubsystemSyntaxPROGram[:SELected]:STATe RUN|PAUSe|STOP|CONTinueExampleOUTPUT 714;"PROGram:SELected:STATe

Strona 344 - The Memory Card Battery

408S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram Subsystem:STRing <varname>{,<svalues>} Th

Strona 345 - Memory Card Initialization

409Chapter 7, IBASIC ControllerPROGram SubsystemSyntaxPROGram[:SELected]:STRing <varname>{,<svalues>}Example setting the value of a simpl

Strona 346

41Chapter 1, Using GPIBGetting Started• Guideline #4. Use single quotes and spaces properly. The syntax diagrams in Chapter 4, “GPIB Commands,” show

Strona 347 - Copying an Entire Volume

410S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemFor simple string variables the value is retur

Strona 348

411Chapter 7, IBASIC ControllerPROGram SubsystemExample of querying the value of a string array whose name is known but whose cur-rent size is unknow

Strona 349 - Using RAM Disk

412S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemCAUTION: The Test Set will continue to process

Strona 350 - Initializing RAM Disks

413Chapter 7, IBASIC ControllerPROGram Subsystem:WAIT? The :WAIT? query command stops the Test Set from executing any commands or queries received t

Strona 351 - Using External Disk Drives

414S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemExample BASIC program without using the :WAIT?

Strona 352

415Chapter 7, IBASIC ControllerPROGram SubsystemEntering a new IBASIC program line IBASIC program lines can be entered directly into the Test Set’s R

Strona 353 - IBASIC Controller

416S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemListing A Program Execute the following comman

Strona 354 - Introduction

417Chapter 7, IBASIC ControllerPROGram SubsystemUploading a Program From the Test Set The following BASIC program copies an IBASIC program from the T

Strona 355 - The IBASIC Controller Screen

418S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerPROGram SubsystemSaving an IBASIC Program To A Memory Card The

Strona 356

419Chapter 7, IBASIC ControllerThe TESTS SubsystemThe TESTS Subsystem The Test Set makes available to the user an automated user-interface which has

Strona 357

42S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting Started• Guideline #6. Numeric values are returned in GPIB Uni

Strona 358 - Program Development

420S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerThe TESTS SubsystemWriting Programs For the TESTS Subsystem The

Strona 359 - Method 3

421Chapter 7, IBASIC ControllerThe TESTS SubsystemProcedure Files A Procedure allows the user to define which of the test subroutines, parameters, an

Strona 360

422S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerThe TESTS SubsystemTESTS Subsystem Screens The TESTS Subsystem

Strona 361

423Chapter 7, IBASIC ControllerThe TESTS SubsystemFigure 30 The TESTS (Main Menu) Subsystem ScreenTESTS Subsystem User-Interface Screens The TESTS S

Strona 362

424S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ibasic.fb Chapter 7, IBASIC ControllerThe TESTS SubsystemTo access any of these screens, position the

Strona 363 - 98642-66505

4258 Programming the Call Processing SubsystemThis chapter presents information on how to control the Test Set’s Call Processing Subsystem using the

Strona 364

426S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemDescription of the Call Processing Su

Strona 365

427Chapter 8, Programming the Call Processing SubsystemDescription of the Call Processing Subsystem’s Remote User InterfaceFor forward control channe

Strona 366 - Receive and Transmit Pacing

428S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemDescription of the Call Processing Su

Strona 367 - PC Configuration

429Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s Remote User InterfaceUsing the Call Processing Subsystem

Strona 368

43Chapter 1, Using GPIBGetting StartedPreparing the Test Set For GPIB Use 1. If other GPIB devices are in the system, attach a GPIB cable from the Te

Strona 369

430S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s

Strona 370

431Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s Remote User InterfaceAccessing the Call Processing Subsy

Strona 371

432S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s

Strona 372 - Terminal Configuration

433Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s Remote User InterfaceConditioning the Test Set for Call

Strona 373

434S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s

Strona 374 - Method 2

435Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s Remote User InterfaceCall Processing Status Register Gro

Strona 375 - Test Set

436S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s

Strona 376

437Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s Remote User InterfaceWhen To Query Data Messages Receive

Strona 377 - Operating System

438S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemUsing the Call Processing Subsystem’s

Strona 378

439Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenProgramming the CALL CONTROL ScreenFigure 34 The CALL CONT

Strona 379

44S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting StartedUsing the GPIB with the Test Set’s built-in IBASIC Cont

Strona 380 - Chapter 7, IBASIC Controller

440S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenAc

Strona 381

441Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenAF FreqThe AF Freq field is displayed only when the Display

Strona 382

442S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenCh

Strona 383

443Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenCntl ChannelThis field is used to set the control channel n

Strona 384

444S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL Screen[]

Strona 385 - (Least Preferred)

445Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenSetting the Display field to DataWhen the Display field is

Strona 386

446S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenIn

Strona 387

447Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenTable 48, “Information Strings Available When A Decoding Er

Strona 388

448S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenRe

Strona 389

449Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenFigure 36 CALL CONTROL Screen with Meas SelectedReading Th

Strona 390

45Chapter 1, Using GPIBGetting StartedBasic Programming Examples The following simple examples illustrate the basic approach to controlling the Test

Strona 391

450S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenES

Strona 392

451Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenFM DeviationThis field displays the measured FM deviation o

Strona 393 - Serial Ports 9 and 10

452S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenHa

Strona 394

453Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenNOTE: The present values for the MS Id fields are:• Phone N

Strona 395 - Serial Port 10 Information

454S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenOr

Strona 396 - PROGram Subsystem

455Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenPageThis field is used to initiate a page to the mobile sta

Strona 397

456S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenPh

Strona 398 - PROGram Subsystem Commands

457Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenPwr Lvl:The Pwr Lvl: field is divided into two fields: • Th

Strona 399

458S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenRe

Strona 400

459Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenReleaseThis field is used to terminate an active voice chan

Strona 401

46S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBGetting StartedThe following example reads several fields. ExampleOUTP

Strona 402

460S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenSA

Strona 403

461Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenSCM:This field displays the decoded station class mark info

Strona 404

462S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenSy

Strona 405

463Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONTROL ScreenTX Freq ErrorThis field displays the frequency error (frequ

Strona 406 - (STATe command sent

464S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL DATA ScreenProgr

Strona 407

465Chapter 8, Programming the Call Processing SubsystemProgramming the CALL DATA Screen[] AccessSee “[] Access” on page 439 for programming informat

Strona 408

466S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL DATA ScreenHando

Strona 409 - IBASIC program

467Chapter 8, Programming the Call Processing SubsystemProgramming the CALL DATA ScreenReading the CALL DATA Screen Message FieldsThis section provid

Strona 410

468S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field Descri

Strona 411

469Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field DescriptionsSThis field displays whether the serial number word

Strona 412 - PROGram[:SELected]:WAIT

47Chapter 1, Using GPIBRemote OperationRemote Operation The Test Set can be operated remotely through the General Purpose Interface Bus (GPIB). Excep

Strona 413

470S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field Descri

Strona 414 - Using the EXECute Command

471Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field DescriptionsORDERThis field displays the Order field and identi

Strona 415

472S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field Descri

Strona 416

473Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field DescriptionsRECCW D Message FieldsFigure 41 RECCW D Message Fi

Strona 417

474S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field Descri

Strona 418

475Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field DescriptionsRVCOrdCon Message FieldsFigure 43 RVCOrdCon Messag

Strona 419 - The TESTS Subsystem

476S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field Descri

Strona 420

477Chapter 8, Programming the Call Processing SubsystemCALL DATA Screen Message Field DescriptionsQuerying Multiple Fields With Single OUTPUT/ENTERWh

Strona 421 - Procedure Files

478S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenProgra

Strona 422 - TESTS Subsystem Screens

479Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenFor example: changing the value of the SAT color code field (SC

Strona 423

48S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBRemote OperationRemote Capabilities Conformance to the IEEE 488.1-1987

Strona 424

480S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenWhen t

Strona 425 - Subsystem

481Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT Screen[] AccessSee “[] Access” on page 439 for programming informati

Strona 426

482S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenThe :D

Strona 427 - Operational Overview

483Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenPageSee “Page” on page 455 for programming information.[] Page

Strona 428

484S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenReadin

Strona 429 - Connecting a Mobile Station

485Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenQuerying Multiple Fields With Single OUTPUT/ENTERWhen multiple

Strona 430

486S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL BIT ScreenModify

Strona 431 - ENTER 714;Screen$

487Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsCALL BIT Screen Message Field DescriptionsThis secti

Strona 432 - Command Syntax

488S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 433

489Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsSPC WORD2 Message FieldsFigure 46 SPC WORD2 Message

Strona 434

49Chapter 1, Using GPIBAddressingAddressing Factory Set Address The Test Set’s GPIB address is set to decimal 14 at the factory. The address can be c

Strona 435

490S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 436

491Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsCPACombined paging/access field.1 binary character r

Strona 437

492S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 438

493Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsBISBusy-Idle status field.1 binary character require

Strona 439 - [] Access

494S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 440 - [] Active

495Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsREGINCRRegistration increment field.12 binary charac

Strona 441 - Called Number:

496S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 442

497Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsOHDThis field displays the overhead message type. •

Strona 443 - Cntl Channel

498S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 444 - [] Connect

499Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsCMACControl mobile attenuation field. Indicates the

Strona 445

5Safety SymbolsCaution, refer to accompanying documentsWarning, risk of electric shockEarth (ground) terminalAlternating currentFrame or chassis ter

Strona 446

50S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBIEEE 488.1 Remote Interface Message CapabilitiesIEEE 488.1 Remote Inte

Strona 447

500S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 448 - ENTER 714;Rcv_data$

501Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsMS WORD1 Message FieldsFigure 51 MS WORD1 Message F

Strona 449

502S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 450 - ESN (hex):

503Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsLOCALThis field is specific to each system. The ORDE

Strona 451 - FM Deviation

504S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 452 - :HANDoff

505Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsVMACThis field displays the voice mobile attenuation

Strona 453

506S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 454 - ENTER 714;Last_ord_sent$

507Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsORDQOrder qualifier field. Qualifies the order to a

Strona 455

508S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field Descrip

Strona 456 - Phone Num:

509Chapter 8, Programming the Call Processing SubsystemCALL BIT Screen Message Field DescriptionsRSVDReserved for future use, all bits must be set as

Strona 457 - Pwr Lvl:

51Chapter 1, Using GPIBIEEE 488.1 Remote Interface Message CapabilitiesClear Lockout/Set LocalYes The Test Set returns to local mode (front-panel con

Strona 458 - [] Register

510S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS ScreenPro

Strona 459 - :RELease

511Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS ScreenRequirements for Using The ANALOG MEAS ScreenThe Test Set mu

Strona 460

512S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS ScreenAFG

Strona 461 - ENTER 714;Sid_number

513Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS ScreenFilter 1This field selects one of several standard or option

Strona 462 - System Type

514S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS ScreenExa

Strona 463 - TX Power

515Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS Screen200 SUB Meas_sinad210 INTEGER Loop_counter220 OUTPUT 714;&q

Strona 464

516S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the ANALOG MEAS ScreenExa

Strona 465 - Display Word

517Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONFIGURE ScreenProgramming the CALL CONFIGURE ScreenFigure 57 The CALL

Strona 466

518S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONFIGURE Screen

Strona 467

519Chapter 8, Programming the Call Processing SubsystemProgramming the CALL CONFIGURE ScreenDetectorThis field is used to select the type of detector

Strona 468 - RECCW A Message Fields

52S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBIEEE 488.1 Remote Interface Message CapabilitiesTrigger Yes If in remo

Strona 469

520S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample ProgramsExample ProgramsThis

Strona 470 - RECCW B Message Fields

521Chapter 8, Programming the Call Processing SubsystemExample ProgramsThe program traps any errors which may occur while executing. If an error is d

Strona 471

522S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample ProgramsSRQ Example Program10

Strona 472 - RECCW C Message Fields

523Chapter 8, Programming the Call Processing SubsystemExample Programs480 BEEP490 DISP "Originate a call from the mobile station."500

Strona 473 - Dig 1 through Dig 8

524S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample Programs3080 ELSE3090

Strona 474 - Dig 9 through Dig 16

525Chapter 8, Programming the Call Processing SubsystemExample Programs6090 CASE "Register"6100 Ptr_value=16110 CASE "Page&

Strona 475 - RVCOrdCon Message Fields

526S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample Programs7200 IF Error_flag

Strona 476

527Chapter 8, Programming the Call Processing SubsystemExample Programs11020 OUTPUT Inst_addr;"DISP ACNT;:CALLP:MOD ’MEAS’;:MEAS:RFR:POW?;FREQ:E

Strona 477 - PRINT Message$

528S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample ProgramsComments for SRQ Exam

Strona 478

529Chapter 8, Programming the Call Processing SubsystemExample Programs2040 * Call Processing Status Register Group Condition register conditions whi

Strona 479

53Chapter 1, Using GPIBRemote/Local ModesRemote/Local Modes Remote Mode In Remote mode all front-panel keys are disabled (except for the LOCAL key, P

Strona 480

530S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample ProgramsPolling Example Progr

Strona 481 - Data Spec

531Chapter 8, Programming the Call Processing SubsystemExample Programs500 Print_error510 END IF520 IF NOT FNSet_state("Release") T

Strona 482 - ENTER 714;Build_method$

532S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample Programs3290 FNEND4000 Set_s

Strona 483 - :MESSage?

533Chapter 8, Programming the Call Processing SubsystemExample Programs6000 Order: DEF FNOrder(Order$,INTEGER Parm)6010 COM /Io_addresses/ INTEGER

Strona 484

534S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample Programs7190 IF BINAND(Std

Strona 485

535Chapter 8, Programming the Call Processing SubsystemExample Programs11190 Timed_out:! 11200 ON TIMEOUT Bus_addr,5 GOTO Cannot_recover11210 CLE

Strona 486

536S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample ProgramsComments for Polling

Strona 487 - SPC WORD1 Message Fields

537Chapter 8, Programming the Call Processing SubsystemExample Programs4060 Ptr_value is the value that the positive transition filter will be set to

Strona 488 - 3 binary characters required

538S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\callproc.fb Chapter 8, Programming the Call Processing SubsystemExample Programs

Strona 490 - 1 binary character required

54S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBRemote/Local ModesLocal To Remote Transitions The Test Set switches fr

Strona 491 - 1 binary characters required

540S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb General Information About Error Messages Information concerning error messages displayed

Strona 492 - ACCESS Message Fields

541Positive Numbered Error Messages Positive numbered error messages are generally associated with IBASIC. Refer to the Instrument BASIC User’s Hand

Strona 493

542S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Negative Numbered Error Messages Negative numbers preceding the error messages text corre

Strona 494 - REG INC Message Fields

543IBASIC Error Messages IBASIC Error Messages are associated with IBASIC operation. IBASIC error messages can have both positive and negative number

Strona 495 - 3 binary character required

544S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb GPIB1 Error Messages GPIB Error Messages are associated with GPIB operation. GPIB error m

Strona 496 - REG ID Message Fields

545Text Only Error Messages Text only error messages are generally associated with manual operation of the Test Set. Refer to the Agilent Technologie

Strona 497

546S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb The Message Display During instrument operation, various messages may appear on the Test

Strona 498 - C-FILMESS Message Fields

547Non-Recoverable Firmware Error The non-recoverable firmware error is very important. It appears when an unanticipated event occurs that the Test S

Strona 499

548S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb GPIB Errors Most GPIB errors occur when the control program attempts to query a measureme

Strona 500

549Text Only GPIB Errors Un-numbered (text only) GPIB error messages are generally self-explanatory. For example, trying to retrieve a saved register

Strona 501 - MS WORD1 Message Fields

55Chapter 1, Using GPIBRemote/Local ModesLocal Lockout The Local Lockout mode disables the front-panel LOCAL key and allows return to Local mode only

Strona 502

550S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb HP-IB Error: Not Enough Memory Available for Save.This message will be generated when the

Strona 503

551Numbered GPIB Error Descriptions The following GPIB errors can be generated under any of the following conditions: • controlling the Test Set with

Strona 504 - MS IntVCh Message Fields

552S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −100 Command errorThis code indicates only that a Command Error as defined in IEEE

Strona 505

553Error −110 Command header errorAn error was detected in the header. Error −111 Header separator errorA character which is not a legal header sepa

Strona 506

554S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −128 Numeric data not allowedA legal numeric data element was received, but the de

Strona 507

555Error −150 String data errorThis error, as well as errors −151 through −158, are generated when parsing a string element. Error −151 Invalid stri

Strona 508 - FVC V Mes Message Fields

556S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −171 Invalid expressionThe expression data element was invalid (see IEEE 488.2, 7.

Strona 509

557Error −202 Settings lost due to rtlIndicates that a setting associated with a hard local control (see IEEE 488.2, 5.6.1.5) was lost when the devi

Strona 510

558S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −221 Settings conflictIndicates that a legal program data element was parsed but c

Strona 511 - AF Anl In

559Error −251 Missing mass storageIndicates that a legal program command or query could not be executed because of missing mass storage. For example

Strona 512 - Detector

56S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\usehpib.fb Chapter 1, Using GPIBRemote/Local Modes

Strona 513 - Filter 2

560S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −260 Expression errorIndicates that an expression program data element related erro

Strona 514 - Example Measurement Routines

561Error −276 Macro recursion errorIndicates that a syntactically legal macro program data sequence could not be executed because the device found i

Strona 515

562S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −286 Program runtime error Error −300 Device-specific errorThis code indicates on

Strona 516

563Error −400 Query errorThis code indicates only that a Query Error as defined in IEEE 488.2 11.5.1.1.7 and 6.3 has occurred. Error −410 Query INTE

Strona 517

564S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −606 Update of Input Module Relay Switch Count file failed.Indicates that the Test

Strona 518 - ENTER 714;Num_acc_chans

565Error −1304 Origination attempted while not in Active state.Indicates that a mobile station attempted to originate a call to the simulated Base S

Strona 519 - TX Pwr Zero

566S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb Error −1311 RF power loss indicates loss of Voice Channel.When any Call Processing Subsy

Strona 520 - Example Programs

567Error −1315 Data from RECC contains invalid bits in word [1,2,3].Indicates that the decoded data received on the reverse control channel contains

Strona 521

568S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\errormsg.fb

Strona 522 - SRQ Example Program

569Symbols*CLS, 220*ESE, 220*ESE?, 220*ESR?, 220*IDN ?, 209*OPC, 213*OPC?, 216*OPT ?, 210*PCB, 221*RCL, 222*RST, 211*SAV, 222*SRE, 221*SRE?, 221*STB

Strona 523

572 Methods For Reading Measurement Results

Strona 524

Index570 BBatterymemory card, 344part numbers, 344replacing, 344CCalibration Status Register Group, 276accessing registers contained in, 274,278condi

Strona 525

Index 571RVCOrdCon, 464ChanCall Control screen, 442clear status, *CLS, 220CMAXCALL CONFIGURE screen, 518Code files, 333, 420Common Commands, 208RST,

Strona 526

Index572 REF SET - setting reference point viaHP-IB, 198REF SET - turning ON/OFF via HP-IB,197turning ON and OFF, 87using AVG via HP-IB, 182using HI

Strona 527

Index 573YES key, 180FVC mobile station control messageorder, 506voice channel assignment, 508HHardware Status Register #1 Group, 284accessing regist

Strona 528

Index574 RESet, 184STATe, 183STATe?, 183VALue, 184VALue?, 184CALLP, 122CONFigure, 117BADDress, 200CPRocess, 122DECoder, 141diagram structure, 92DISPl

Strona 529

Index 575II/O ConfigureHP-IB command syntax diagram, 117IBASICavoiding program hangs, 40command line, 356Controller, 26, 28, 29, 44Controller - defau

Strona 530 - Polling Example Program

Index576 LLibrary files, 420backing up, 346LIF file names, 334LIF file system, 334initializing media for, 338lock upHP-IB bus, 226, 231MManual Contro

Strona 531

Index 577NNon-Recoverable Firmware Error, 547Number MeasurementHP-IB command syntax diagram, 177OOperating Modesexternal automatic control, 26, 30int

Strona 532

Index578 QQuestionable Data/Signal RegisterGroup, 266accessing registers contained in, 269condition register bit assignments, 268RRadio InterfaceHP-I

Strona 533

Index 579standard event status register query,ESR?, 220StatusHP-IB command syntax diagram, 168,169status byte query, *STB?, 221Status Byte Register,

Strona 534

58S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb Chapter 2, Methods For Reading Measurement ResultsBackgroundBackgroundOne of the most comm

Strona 535

Index580 file names (see also DOS & LIF filenames), 335TriggerHP-IB command syntax diagram, 173HP-IB commands, 228Trigger - aborting, 228Trigger

Strona 536

Index 581Wwait to complete, *WAI, 219Wildcards, 348wordabbreviated address, 468extended address, 470first word of called address, 473reverse voice ch

Strona 537

59Chapter 2, Methods For Reading Measurement ResultsBackgroundThis situation can be avoided entirely by:1. sending a Selected Device Clear (SDC) inte

Strona 538

6 Agilent Technologies Warranty Statement for Commercial Products Agilent Technologies 8920A RF Communications Test SetDuration of Warranty: 1 year 1

Strona 539 - Error Messages

60S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘ON TIMEOUT’ Example ProgramHP

Strona 540

61Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘ON TIMEOUT’ Example Program10 COM /Io_names/ INTEGER Inst_addr,Bus_addr20 CLEAR SCREE

Strona 541

62S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘ON TIMEOUT’ Example ProgramCo

Strona 542

63Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘ON TIMEOUT’ Example Program360Exit the Measure function and return the result value.37

Strona 543 - IBASIC Error Messages

64S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘MAV’ Example ProgramHP® BASIC

Strona 544 - Error Messages

65Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘MAV’ Example Program10 COM /Io_names/ INTEGER Inst_addr,Bus_addr20 CLEAR SCREEN30 Ins

Strona 545 - Text Only Error Messages

66S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘MAV’ Example ProgramComments

Strona 546 - The Message Display

67Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘MAV’ Example Program360Perform a serial poll to read the Status Byte from the Test Set

Strona 547

68S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\measrslt.fb Chapter 2, Methods For Reading Measurement ResultsHP® BASIC ‘MAV’ Example Program480Exit t

Strona 548 - GPIB Errors

693 GPIB1 Command Guidelines 1. GPIB was formerly called HP-IB for Hewlett-Packard instruments. Some labels on the instrument may still reflect the f

Strona 549 - Text Only GPIB Errors

78. Agilent Technologies will be liable for damage to tangible property per incident up to the greater of $300,000 or the actual amount paid for the

Strona 550

70S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesSequential and Overlapped CommandsSequential and Overlap

Strona 551

71Chapter 3, GPIB Command GuidelinesGuidelines for OperationGuidelines for Operation The following topics discuss rules and guidelines for controllin

Strona 552

72S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationCommand Punctuation NOTE: Progra

Strona 553

73Chapter 3, GPIB Command GuidelinesGuidelines for OperationUsing Colons to Separate Commands The GPIB command syntax is structured using a control h

Strona 554

74S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationUsing the Semicolon and Colon to

Strona 555

75Chapter 3, GPIB Command GuidelinesGuidelines for OperationSpecifying Units-of-Measure for Settings and Measurement Results Numeric settings and mea

Strona 556

76S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationChanging Display Units. Use th

Strona 557

77Chapter 3, GPIB Command GuidelinesGuidelines for OperationReading Back Display Units Setting. Use the Display Units query command, DUNits?, to re

Strona 558

78S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationGPIB Units (UNITs) GPIB Units ar

Strona 559

79Chapter 3, GPIB Command GuidelinesGuidelines for OperationChanging GPIB Units. Use the UNITs command to change the GPIB Units setting for selecte

Strona 560

8 DECLARATION OF CONFORMITY according to ISO/IEC Guide 22 and EN 45014Manufacturer’s Name:Manufacturer’s Address:declares

Strona 561

80S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationReading-Back GPIB Units. Use t

Strona 562

81Chapter 3, GPIB Command GuidelinesGuidelines for OperationAttribute Units (AUNits) Attribute Units are the units-of-measure used by the Test Set wh

Strona 563

82S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationFor each measurement that can be

Strona 564

83Chapter 3, GPIB Command GuidelinesGuidelines for OperationChanging Attribute Units. The AUNits command can be used to change the Attribute Units

Strona 565

84S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationIf it is not possible to properl

Strona 566

85Chapter 3, GPIB Command GuidelinesGuidelines for OperationDisplay Units and GPIB Units are not affected when changing Attribute Units. !Reset the T

Strona 567

86S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationReading-back Attribute Units.

Strona 568

87Chapter 3, GPIB Command GuidelinesGuidelines for OperationUsing the STATe Command The STATe command corresponds to the front-panel ON/OFF key and i

Strona 569

88S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for OperationReading back the measurement, in

Strona 570

89Chapter 3, GPIB Command GuidelinesGuidelines for OperationSample GPIB Program The following program was written on an HP® 9000 Series 300 controlle

Strona 571

9Table 1 Regional Sales OfficesUnited States of America:Agilent TechnologiesTest and Measurement Call Center P.O. Box 4026Englewood, CO 80155-4026(t

Strona 572

90S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\guidelin.fb Chapter 3, GPIB Command GuidelinesGuidelines for Operation380 ! Read the oscilloscope trac

Strona 573

914 GPIB Commands

Strona 574

92S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ch3intro.fb Chapter 4, GPIB CommandsGPIB Syntax DiagramsGPIB Syntax DiagramsGPIB Command Syntax Diagra

Strona 575

93Chapter 4, GPIB CommandsGPIB Syntax DiagramsInstrument Command Number Setting Syntax DiagramsInteger Number Setting Syntax, page 174.Real Number Se

Strona 576

94S:\agilent\8920\8920b\PRGGUIDE\BOOK\CHAPTERS\ch3intro.fb Chapter 4, GPIB CommandsGPIB Syntax DiagramsDiagram ConventionsUse the following diagram t

Strona 577

95Adjacent Channel Power (ACP)Adjacent Channel Power (ACP) :ACPower’spaceRatioLevelReturns quoted string?’:MEASurement:CBANSee Real Number Setting Sy

Strona 578

96S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\acp.sec Adjacent Channel Power (ACP)

Strona 579

97AF AnalyzerAF Analyzer ’spaceGndFloat600 To HiReturns quoted string?’:AFANalyzer:AIN’space0 dB10 dB20 dB30 dBReturns quoted string?’:CURRent’space7

Strona 580

98S:\agilent\8920\8920b\PRGGUIDE\BOOK\SECTIONS\afan.sec AF Analyzer’space<20Hz HPF50Hz HPF300Hz HPFOptional FiltersReturns quoted string?’:AFANaly

Strona 581

99AF Analyzer:AFANalyzer:NOTCh:FREQuencySee Real Number Setting Syntax**Does not include the :STATe command’space0 dB10 dB20 dB30 dB40 dBReturns quot