Agilent Technologies 85032B/E 50 Instrukcja Użytkownika Strona 1

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Pomiary, testowanie i kontrola Agilent Technologies 85032B/E 50. Agilent 8712B and 8714B RF Economy Vector Network Analyzers Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 12
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 0
REFLECTION
TRANSMISSION
Agilent 8712B and 8714B
RF Economy Vector
Network Analyzers
Data Sheet
8712B, 300kHz to 1.3 GHz
8714B, 300kHz to 3.0 GHz
Przeglądanie stron 0
1 2 3 4 5 6 ... 11 12

Podsumowanie treści

Strona 1 - Network Analyzers

REFLECTIONTRANSMISSIONAgilent 8712B and 8714BRF Economy VectorNetwork AnalyzersData Sheet8712B, 300kHz to 1.3 GHz8714B, 300kHz to 3.0 GHz

Strona 2 - Maximum and Minimum Power

Front panel connectorsConnector type type-N female Impedance 50 ohms (standard) 75 ohms (Option 1EC) Probe power +15V 200 mA -12.6V 250 mA Rear panel

Strona 3 - Receiver

11This document describes the system performance ofthe 8712B and 8714B 50 ohm and 75 ohm (Option1EC) network analyzers, and provides two kinds ofinfor

Strona 4 - Group Delay

Agilent Technologies’ Test and Measurement Support, Services, and AssistanceAgilent Technologies aims to maximize the value you receive,while minimizi

Strona 5 - Supplemental Data

SpecificationsMeasurement ports 8712B 8714B50 and 75 ohmDirectivity 40 dB 40 dBSource match 20 dB 20 dB(reflection)Source match 14 dB typical123 dB ty

Strona 6 - Characteristics

3Receiver Frequency range 8712B 8714BNarrowband 300 kHz 300 kHzto 1.3 GHz to 3.0 GHzBroadband 0.10 to 1.3 GHz 0.10 to 3.0 GHzDynamic range2Narrowband

Strona 7

4AM Delay (Option 1DA/1DB)This option adds amplitude modulation groupdelay capability, which allows measurements ofgroup delay through frequency-trans

Strona 8 - Calibration

5Source Signal PurityNonharmonic spurious Agilent 8712B Agilent 8714B ≥50 kHz from carrier <–20 dBc, <1 MHz <–30 dBc<–30 dBc, ≥1 MHz&

Strona 9 - Special Options

6MeasurementNumber of display channelsTwo display channels are available.Measurements• Narrowband: reflection (A/R), transmission(B/R), A, B, R • Bro

Strona 10 - General Characteristics

7Determining Optional Sweep Speed andDynamic RangeDynamic range, sweep time and IF Bandwidth areinterdependent quantities. Reducing sweep timeusually

Strona 11

8Measurement CalibrationCalibration significantly reduces measurementuncertainty due to errors caused by system direc-tivity, source match, reflection

Strona 12 - 5964-0112E

9Standard Options75 ohms (Option 1EC) Provides 75 ohm system impedance.AM delay (Option 1DA/1DB)This option adds amplitude modulation groupdelay capab

Komentarze do niniejszej Instrukcji

Brak uwag