SCPI Commands Reference SOURce Commands
62 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
[:SOURce]:VOLTage[0|1|2][:LEVel]
[:IMMediate]:HIGH
Syntax
[:SOUR]:VOLT[0|1|2][:LEV][:IMM]:HIGH[?]
Form Set & Query
Description Programs the high- level of the output signal for the trigger output
and the channels.
Parameter Numeric [uV|mV|V] |MIN|MAX
Value coupling Amplitude = High – Low
Offset = (High – Low) / 2
Range coupling Low- level
*RST value Trigger channel (0): 0.1 V
Channels 1 and 2: 0.1 V
Specified limits Trigger channel (0): –1.95 … +3.0 V
Channels 1 and 2: –1.95 ... +3.0 V
Absolute limits Trigger output (0): –2.2 … +3.2 V
Channels 1 and 2: –2.2 … +3.2 V
Example Set Channel 1 high- level to –200 mV.
:VOLT1:HIGH -200MV