SCPI Commands Reference SOURce Commands
58 Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007
NOTE This is not the same as the
:DIGital[1|2][:STIMulus]:SIGNal:POLarity
command, which logically
inverts the 32- bit data on the channels by swapping 1s with 0s
and vice- versa.
[:SOURce][:PULSe]:WIDTh[1|2]
Syntax
[:SOUR][:PULS]:WIDT[1|2][?]
Form Set & Query
Description Programs the pulse width for a channel.
Parameter Numeric [ps|ns|us|ms|s]|MIN|MAX
Functional coupling Programming the pulse width also executes
[:SOURce][:PULSe]:HOLD WIDTh
so that the pulse width is held
constant when the signal frequency is changed.
Value coupling Dutycycle = (Width / Period) * 100
Range coupling Frequency, Period
*RST value 50% of Period
Specified limits 100E–12 … (Period – 100E–12) sec
Absolute limits 100 ps <= <value> <= <Period value> –100 ps && <value> < 10 ns
Example Set Channel 1 pulse width to 1 ns.
:WIDT1 1NS