626 Chapter7
Language Reference
SWPCPL Sweep Couple
SWPCPL Sweep Couple
Syntax
Figure 7-270 SWPCPL Syntax
Description
The SWPCPL command selects either a stimulus-response (SR) or
spectrum-analyzer (SA) auto-coupled sweep time. In stimulus-response
mode, auto-coupled sweep times are usually much faster for
swept-response measurements. Stimulus-response auto-coupled sweep
times are typically valid in stimulus-response measurements when the
system frequency span is less than 20 times the bandwidth of the
device under test.
Preset State
SA (spectrum analyzer mode)
Query Response
Figure 7-271 SWPCPL Query Response