Agilent Technologies 54503A Podręcznik Użytkownika Strona 604

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 716
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 603
604 Chapter7
Language Reference
SRCPOFS Source Power Offset
SRCPOFS Source Power Offset
Syntax
Figure 7-251 SRCPOFS Syntax
Description
The SRCPOFS command offsets the displayed power of the built-in
tracking generator. This function can be used to take into account
system losses (for example, cable loss) or gains (for example,
preamplifier gain) reflecting the actual power delivered to the device
under test. SRCPOFS is available only with an 8560E/EC Option 002.
Parameters
number real from 100 dB to +100 dB.
UP or DN increments in steps equal to the value set by SRCPSTP.
Preset State
0.0 dB
Przeglądanie stron 603
1 2 ... 599 600 601 602 603 604 605 606 607 608 609 ... 715 716

Komentarze do niniejszej Instrukcji

Brak uwag