Agilent Technologies 54502A Instrukcja Obsługi Strona 42

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 111
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 41
c. Move cursor to the bit assignment field and turn on appropriate eight
bits to be tested (* = on; . = off) as in figure below.
5. Set up the State Trace Specification without sequencing levels and set
Count States as in next figure.
a. Press front-panel TRACE key.
b. Move cursor to Count, press SELECT, then move cursor to States,
press SELECT, and set to any state by pressing SELECT again.
Figure 3-7. Bit Assignment for Data Test 1
Figure 3-8. Trace Specification for Data Test 1
HP 1650B/1651B Performance Tests
Service Manual 3-7
Przeglądanie stron 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 110 111