Making Measurements
Using Ripple Limits to Test a Device
Figure 1-65 Connections for an Example Ripple Test Measurement
2. Press and choose the measurement settings. For this example, the
measurement settings are as follows:
or on ET models:
You may also want to select settings for the number of data points, power, averaging,
and IF bandwidth.
3. Substitute a thru for the device and perform a response calibration by pressing:
4. Reconnect your test device.
5. To better view the measurement trace, press . Refer to
Figure 1-66.
Preset
Meas
Trans: FWD S21 (B/R)
TRANSMISSN
Center 1.8 G/n
Span 3.4 G/n
Scale Ref
AUTO SCALE
Cal
CALIBRATE MENU
RESPONSE
THRU
Scale Ref
AUTO SCALE