Specifications
Agilent 81101A Specifications
Leading and trailing edges can be programmed independently within the
following ranges (Maximum ratio 1:20):
Level Specifications
External Load compensation
For loads
50
, the actual load impedance can be entered to correct the
output values.
Level Parameters Agilent 81101A
Source impedance: selectable 50
± 1% typical or 1 k
Maximum external voltage: ±24 V
Short circuit current: ±400 mA
Normal/complement: selectable
ON/OFF: relay connect/disconnect output (HiZ).
Limits: high and low levels can be limited to protect the
DUT
5ns