Agilent Technologies N1022A Podręcznik Użytkownika Strona 15

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 18
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 14
15
Calibration Test Record
Calibration Test Record
Agilent Technologies
N2790A 100 MHz Differential Probe
Serial No.:_______________________
Certification Date:_________________
Tested By:_______________________
Recommended Test Interval: 1 Year
Recommended Date of Next Certification:_________
Certification Temperature:_____________________
_______________________________
_______________________________
Test Probe Settings Limits Results
Gain 50:1 +196 mV ... +204 mV (\2% of reading)
Gain 50:1 +1.96 V ... +2.04 V (\2% of reading)
Gain 500:1 +196 mV ... +204 mV (\2% of reading)
Gain 500:1 +1.96 V ... +2.04 V (\2% of reading)
Gain 50:1 -196 mV ... -204 mV (\2% of reading)
Gain 50:1 -1.96 V ... -2.04 V (\2% of reading)
Gain 500:1 -196 mV ... -204 mV (\2% of reading)
Gain 500:1 -1.96 V ... -2.04 V (\2% of reading)
Bandwidth 50:1 > 2.8 div Reading on oscilloscope
Bandwidth 500:1 > 2.8 div Reading on oscilloscope
Przeglądanie stron 14
1 2 ... 10 11 12 13 14 15 16 17 18

Komentarze do niniejszej Instrukcji

Brak uwag