Output Configuration Commands 4
U2761A Programmer’s Reference 47
[SOURce:]VOLTage[:LEVel][:IMMediate]:OFFSet
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate]:OFFSet <offset>
This command sets the offset voltage.
[SOURce:]VOLTage[:LEVel][:IMMediate]:OFFSet?
This query returns the offset voltage for the currently selected function.
Parameter
Examples
The following command sets the offset voltage to 1 V.
VOLT:OFFS 1
The following query returns the offset voltage.
VOLT:OFFS?
Typical response:
+1.0000000000000E+00
Item Type Range of values Default value
offset Numeric ±2.48 V (Into 50 Ω load)
[1]
±4.96 V (Into open circuit)
[1]
[1] For the DC function, the range of values are ±2.5 V (into 50 Ω load) and ±5 V (into open circuit).
0 V