Agilent Technologies HP 3325B Instrukcja Obsługi Strona 72

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 71
Chapter 1 Performance Tests
Test 1: Trigger Level (Agilent 53181A Only)
1-48 Assembly-Level Service Guide
1
Test 1: Trigger Level (Agilent 53181A Only)
This test verifies the Trigger Level accuracy of the Agilent 53181A Frequency
Counter.
Equipment
Agilent 3325B Synthesizer
Agilent 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set CHANNEL 1 input condition to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000V
50/1M, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
Przeglądanie stron 71
1 2 ... 67 68 69 70 71 72 73 74 75 76 77 ... 251 252

Komentarze do niniejszej Instrukcji

Brak uwag