Agilent Technologies HP 3325B Instrukcja Obsługi Strona 67

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 66
Chapter 1 Performance Tests
Agilent 53131A/132A Performance Test Record (Page 3 of 6)
Assembly-Level Service Guide 1-43
1
Agilent 53131A/132A Performance Test Record
(Page 3 of 6)
Complete Performance Tests (Continued)
Test
Number Test Description Minimum
Actual Reading
Maximum
4 Channels 1 and 2
Frequency Accuracy
(Agilent 53131A/132A
Only)
1 MHz Test
(Agilent 53131A):
Channel 1
Channel 2
1 MHz Test
(Agilent 53132A):
Channel 1
Channel 2
999,999.999 Hz
999,999.999 Hz
999,999.9998 Hz
999,999.9998 Hz
1. __________
2. __________
3. __________
4. __________
1,000,000.001 Hz
1,000,000.001 Hz
1,000,000.0002 Hz
1,000,000.0002 Hz
4 100 MHz Test
(Agilent 53131A):
Channel 1
Channel 2
100 MHz Test
(Agilent 53132A):
Channel 1
Channel 2
99.999,999,93 MHz
99.999,999,93 MHz
99.999,999,980 MHz
99.999,999,980 MHz
5. __________
6. __________
7. __________
8. __________
100.000,000,07 MHz
100.000,000,07 MHz
100.000,000,020 MHz
100.000,000,020 MHz
Przeglądanie stron 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 251 252

Komentarze do niniejszej Instrukcji

Brak uwag