Agilent Technologies 6633A Podręcznik Użytkownika Strona 99

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 176
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 98
Language Dictionary - 8
99
MEASure:CURRent:LOW?
FETCh:CURRent:LOW?
Agilent 66311B Only
These queries return the Low level current of a current pulse waveform. The instrument first measures
the minimum and maximum data points of the pulse waveform. It then generates a histogram of the pulse
waveform using 1024 bins between the maximum and minimum data points. The bin containing the most
data points below the 50% point is the low bin. The average of all the data points in the low bin is
returned as the Low level. If no low bin contains more than 1.25% of the total number of acquired points,
then the minimum value is returned by these queries.
Query Syntax
MEASure[:SCALar]:CURRent:LOW?
FETCh[:SCALar]:CURRent:LOW?
Parameters
None
Examples
MEAS:CURR:LOW? FETC:CURR:LOW?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:HIGH? CALC:REF:LOW
MEASure:CURRent:MAXimum?
FETCh:CURRent: MAXimum?
Agilent 66311B Only
These queries return the maximum output current.
Query Syntax
MEASure[:SCALar]:CURRent:MAXimum?
FETCh[:SCALar]:CURRent:MAXimum?
Parameters
None
Examples
MEAS:CURR:MAX? FETC:CURR:MAX?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:MIN?
MEASure:CURRent:MINimum?
FETCh:CURRent:MINimum?
Agilent 66311B Only
These queries return the minimum output current.
Query Syntax
MEASure[:SCALar]:CURRent:MINimum?
FETCh[:SCALar]:CURRent:MINimum?
Parameters
None
Examples
MEAS:CURR:MIN? FETC:CURR:MIN?
Returned Parameters
<NR3>
Related Commands
MEAS:CURR:MAX?
Przeglądanie stron 98