TRIGger:SEQuence2:HYSTeresis:CURRent
TRIGger:ACQuire:HYSTeresis:CURRent
This command defines a band around the trigger level through which the signal must pass before an
internal measurement can occur. The band limit above and below the trigger level is one half of the
hysteresis value added to or subtracted from the trigger level.
For a positive trigger to occur, the excursion of an output waveform in the positive direction must start
below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative
trigger to occur, the excursion of an output waveform in the negative direction must start above the upper
hysteresis band limit and pass through the lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:HYSTeresis:CURRent<NRf+>
TRIGger:ACQuire:HYSTeresis:CURRent<NRf+>
Parameters
0 to MAX (see table 8-3)
Unit
A (amperes)
*RST Value
0
Examples
TRIG:SEQ2:HYST:CURR 0.5
TRIG:ACQ:HYST:CURR 0.5
Query Syntax
TRIGger:SEQuence2:HYSTeresis:CURRent?
TRIGger:ACQuire:HYSTeresis:CURRent?
Returned Parameters
<NR3>
Related Commands
TRIG:SEQ2:HYST:VOLT TRIG:SEQ2:LEV:CURR
TRIGger:SEQuence2:HYSTeresis:VOLTage
TRIGger:ACQuire:HYSTeresis:VOLTage
This command defines a band around the trigger level through which the signal must pass before an
internal measurement can occur. The band limit above and below the trigger level is one half of the
hysteresis value added to or subtracted from the trigger level.
For a positive trigger to occur, the excursion of an output waveform in the positive direction must start
below the lower hysteresis band limit and pass through the upper hysteresis band limit. For a negative
trigger to occur, the excursion of an output waveform in the negative direction must start above the upper
hysteresis band limit and pass through the lower hysteresis band limit.
Command Syntax
TRIGger:SEQuence2:HYSTeresis:VOLTage<NRf+>
TRIGger:ACQuire:HYSTeresis:VOLTage<NRf+>
Parameters
0 to MAX (see table 8-3)
Unit
V (volts)
*RST Value
0
Examples
TRIG:SEQ2:HYST:VOLT 2
TRIG:ACQ:HYST:VOLT 2
Query Syntax
TRIGger:SEQuence2:HYSTeresis:VOLTage?
TRIGger:ACQuire:HYSTeresis:VOLTage?
Returned Parameters
<NR3>