Module 12
Miscellaneous Operations
12-11
Advanced Options for CMU Calibration
Reference Standard Values area provides the following entry fields used to set the reference values
of the open/short/load standard. If you use the standard, enter the reference values to the entry fields.
OPEN Reference: Capacitance and Conductance. If you do not use the reference, enter 0 in the both
fields.
SHORT Reference: Inductance and Resistance. If you do not use the reference, enter 0 in the both
fields.
LOAD Reference: Inductance and Resistance