Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
31
Connections for 50 and 62 GHz oscilloscope models, channels 1R- 3R
Procedure
1
Preset the power meter.
2 Ensure that the power sensor is disconnected from any source and zero the meter.
3 Connect the power sensor to the power meter's Power Ref connector and calibrate the
meter.
4 Make the connections to oscilloscope channel 1 as shown in the connection diagram
above.
5 Set up the Power Meter to display measurements in units of Watts.
6 Press [Default Setup], then configure the scope as follows:
a Ensure Channel 1 is displayed and all other channels are turned off.
b Set the vertical scale of channel 1 to 10 mV/div.
c Set the horizontal scale to 16 ns/div (to display 8 cycles of a 50 MHz waveform).
Notes
Connect output 1 of the V241C splitter to the scope Channel n input directly without any additional
cabling or adapters.
Minimize the use of any adapters.
Ensure that 1.85 mm connectors are tightened properly:
8 in-lbs (90 N-cm) for 1.85 mm
Power sensor
cable
Power sensor
N8488A
Power splitter
Anritsu V241C
Microwave cable
Gore 67 GHz
Microwave CW
Generator E8257D
Power meter
N1912A
Oscilloscope
Click here and enter 16 ns