Chapter 3: Testing Performance
Offset Accuracy Test
17
b When the Acquisition Setup window appears, enable averaging and set the # of
averages to 256 as shown below.
3 Configure the scope to measure Average voltage on channel 1 as follows:
a Change the vertical scale of channel 1 to 10 mV/div.
b Click the V avg measurement on the left side of the screen.
V avg
measurement