Agilent Technologies E1356A Dokumentacja Strona 61

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 149
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 60
Poisson 1/2 Bridge
(HPOisson)
In applications where there is no equal strain of opposite sign as there is for
the beam in Figure 4-10, an increase in measurement sensitivity and
temperature compensation is achieved by mounting the strain gages at right
angles on the specimen. Mounting the gages in this manner is known as a
Poisson arrangement and is represented in Figure 4-11. With a Poisson
arrangement, the sensitivity is not as great as with the bending
1
2
bridge;
however, it is approximately 25% greater than a
1
4
bridge arrangement.
Again, since the gages are adjacent legs of the Wheatstone Bridge circuit,
complete temperature compensation is achieved.
Dummy Gage
In a dummy gage configuration, one of the strain gages in the
1
2
bridge
arrangement is mounted to an unstrained portion of the specimen (Figure
4-12). This gage arrangement provides complete temperature
compensation; however, measurement sensitivity is not increased since a
1
4
bridge measurement is made with the one active gage.
Figure 4-12. Dummy Gage Arrangement
Figure 4-11. Strain Gages in Poisson Arrangement
60 Understanding the Strain Gage Multiplexers Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Przeglądanie stron 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 148 149

Komentarze do niniejszej Instrukcji

Brak uwag