Appendix E
Strain Gage Multiplexer Error Messages
Table E-1 lists the error messages associated with the strain gage
multiplexer modules programmed by SCPI. See mainframe manual for a
complete list.
Table E-1. Strain Gage Multiplexer Error Messages
No. Title Potential Cause(s)
-211 Trigger ignored Trigger received when scan not enabled. Trigger received after scan
complete. Trigger too fast.
-213 Init ignored Attempting to execute an INIT command when a scan is already in
progress.
-224 Illegal parameter value Attempting to execute a command with a parameter not applicable to the
command.
+1500 External trigger source
already allocated
Assigning an external trigger source to a switchbox when the trigger source
has already been assigned to another switchbox.
+2000 Invalid card number Addressing a module (card) in a switchbox that is not part of the switchbox.
+2001 Invalid channel number Attempting to address a channel of a module in a switchbox that is not
supported by the module (e.g., channel 99 of a multiplexer module).
+2006 Command not
supported
Sending a command to a module (card) in a switchbox that is unsupported
by the module.
+2008 Scan list not initialized Executing INIT command with no scan <channel_list>.
+2009 Too many channels in
channel list
Attempting to address more channels than available in the switchbox.
+ 2010 Scan mode not allowed
on this card
Attempting to execute a SCAN:MODE which is not supported.
+ 2011 Empty channel list No valid channels in channel list.
+2012 Invalid channel range Invalid channel(s) specified in SCAN <channel_list> command. Attempting to
begin scanning when no valid channel list is defined.
+ 2017 Incorrect ARM:COUNt Attempting to run a downloaded scan list with ARM:COUNt set to value
other than 1.
+2600 Function not supported
on this card
Sending a command to a module (card) in a switchbox that is not supported
by the module or switchbox.
+2601 Channel list required Sending a command requiring a channel list without the channel list.