Agilent N5161A/62A/81A/82A/83M MXG Signal Generators User’s Guide 29
Setting Preferences & Enabling Options
Upgrading Firmware
back. In this case, you can re- enable the signal generator’s ability to use time–based licenses by
moving the clock forward to the original time or simply waiting that length of time.
Reference Oscillator Tune
Upgrading Firmware
For information on new firmware releases, go to http://www.agilent.com/find/upgradeassistant.
Tunes the internal VCTXCO oscillator frequency.
The user value offsets the factory tuned value (the value is added to the
factory calibrated DAC value). The tune value of 0 sets the factory
calibrated value.
The range; –8192 to 8192, can be set by using the front panel keypad,
knob or remote command.
For details on each key, use key help
as described on page 42.
See also the SCPI Command Reference.