Agilent Technologies E5500 Phase Noise Measurement System
Evaluating Your Measurement Results
Evaluating the Results
For example, applying the 7 dB difference in noise levels, shown in
“Example Comparison of Measurement Results and Reference Source
Noise.” on page 15-4 at 10 kHz, to the graph, reveals that the measured
results should be decreased by about 1 dB at 10 kHz to reflect the actual
noise of the UUT.
Figure 15-2 Example Comparison of Measurement Results and Reference Source
Noise.
Przeglądanie stron 362