Agilent E5500 A-Series
Preconfigured Solutions
The E5500 A-series solutions have
been configured to meet the needs of
production test. They include industry
standard VXI and MMS modular
hardware to provide maximum test
throughput, current and future meas-
urement flexibility, and asset reuse.
These solutions can easily include
specific RF reference sources, addi-
tional baseband analysis hardware,
and computers for complete production
phase noise measurement solutions.
E5501A 50 kHz to 1.6 GHz
E5502A 50 kHz to 6 GHz
E5503A 50 kHz to 18 GHz
E5504A 50 kHz to 26.5 GHz
The Agilent E5500 A-series solution
hardware includes:
• 0.01 Hz to 4 MHz VXI digitizer
• VXI to PC high speed interface
• 6-slot, high power VXI mainframe
• MMS baseband test set and frame
E5502/03/04A solutions add an MMS
low noise downconverter.
Optional Capabilities
• Extend offset range to 8, 10, and
100 MHz
• Add RF reference source
• Add high power input capability
(includes µW phase and AM
detectors)
• Add specific ATE computer
• Extend carrier frequency to
110 GHz
Operating Characteristics
Offset frequency range:
0.01 Hz to 4 MHz
System noise response:
1
–180 dBc/Hz typically
(>10 kHz offsets)
System spurious response:
1
<–120 dBc typically
Phase detector input power:
(<1.6 GHz carrier frequency)
R input = 0 to +23 dBm
L input = +15 to +23 dBm
Downconverter input range:
1 GHz to 6 GHz
1 GHz to 18 GHz
1.5 GHz to 26.5 GHz
External noise input port:
0.01 Hz to 100 MHz
Measurement accuracy:
±2 dB (<1.0 MHz offsets)
±4 dB (<100 MHz offsets)
1. Without reference sources or downconverters.