Agilent PXT Wireless Communications Test Set
User’s Guide
116
Short DRX Cycle
Connected Mode DRX related – present in RRC Connection Reconfiguration. Controls whether shortDRX IE
present or not.
Overwrites Scenario File Value
Dependencies and/or Couplings
Connected DRx On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More
Short DRx-Cycle
Specifies the number of subframes between the onDurationTimer starting for each short cycle
Specifies the number of consecutive subframe(s) for which the UE follows the Short DRX cycle. Value sf2
corresponds to 2 sub-frames, sf5 corresponds to 5 subframes and so on.
Overwrites Scenario File Value
Dependencies and/or Couplings
Short DRX Cycle On/Off must be set to On.
BSE > Mode Setup> More > RRC Settings > More
RRC – Key Menu 3
Accesses the menu that enables you to configure the following settings:
drxShortCycleTimer
Timers
Additional Spectrum Emission
SIB1
drxShortCycleTimer
Specifies the Timer for the DRX Short Cycle as multiples of the shortDRX-Cycle. For example, entering a
value of 3 for this setting corresponds to 3 times the number of subframes you specified for shortDRX-
Cycle.
Overwrites Scenario File Value