Agilent Technologies 8510C Instrukcja Obsługi Strona 131

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 130
8510XF Network Analyzer Systems
4-13
Performance Verification
Verification Procedures
12. When verification has been completed view the data, then print or save
(if required). If needed the verification standard can be remeasured by
selecting:
Repeat Measure.
13. Select
Select Standard to return to the Standard Selection Menu. Repeat
steps 7 through 12 above for the other verification standard (refer to
Table 4-6 on page 4-10).
14. Select
Prior Menu until the Main Menu appears.
15. Select
Quit Program to exit the program, then close the program window to
finish.
Przeglądanie stron 130
1 2 ... 126 127 128 129 130 131 132 133 134 135 136 ... 221 222

Komentarze do niniejszej Instrukcji

Brak uwag