118 Agilent 34980A Mainframe User’s Guide
3 Features and Functions
The sweep count is valid only while scanning. If no channels have been
assigned to the scan list, the specified sweep count is ignored (no error
is generated).
You can specify a sweep count in conjunction with a trigger count and
a sample count. The three parameters operate independent of one
another, and the total number of readings returned will be the product
of the three parameters.
You can store at least 500,000 readings in memory and all readings are
automatically time stamped. If memory overflows, the new readings will
overwrite the first (oldest) readings stored; the most recent readings are
always preserved.
The
CONFigure and MEASure? commands automatically set the sweep
count to 1 sweep.
The instrument sets the sweep count to 1 after a Factory Reset
(
*RST command). An Instrument Preset (SYSTem:PRESet command) or
Card Reset (
SYSTem:CPON command) does not change the setting.
Front Panel Operation: Scan (Configure) > SWEEP COUNT
Remote Interface Operation: SWEep:COUNt
Sample Count
The sample count sets the number of auto- triggered samples the internal
DMM will take per channel per trigger. The sample count applies to both
scanning and stand- alone DMM measurements (with no scan list). The
front- panel sample annunciator (“ * ”) turns on during each measurement.
Sample count for Stand-Alone DMM Mode
t
Sample Count
(1 to 500,000 samples)