146 N9320A Spectrum Analyzer
6 Instrument Messages
215 TG start freq is less than 1/2 res bw
Tracking generator uncalibrated at start frequencies below
1/2 the current resolution bandwidth.
224 Option not licensed
The selected option requires a license. Refer to the
installation procedures in the user guide provided for this
particular option.
238 VBW filter or Average Detector cannot use both
501 RF over load
Either increase input attenuation or decrease the input
signal level.
502 IF over load
Either increase input attenuation or decrease the input
signal level.
604 File already exists
You attempted to save to a file that already exists. Either
delete or rename the existing file or select a new name, then
try again.
605 Media is protected
A save was attempted to a write- protected device.
610 File access is denied
The file is protected or hidden and cannot be accessed.
612 File not found
The analyzer could not find the specified file.
651 Connect TG OUT to RF IN
Attempt to align the tracking generator without its output
connected. Connect the TG SOURCE to RF IN.