202 Chapter 4
Calibration
Specifying Different Standard for Each Frequency
Specifying Different Standard for Each Frequency
Defining different standard for each frequency band
This section demonstrates the procedure to define a different open standard for each
frequency band, based on the following information.
Procedure
Step 1. Press .
Step 2. Press
Cal Kit.
Step 3. Select a calibration kit you want to define.
Step 4. Press
Modify Kit.
Step 5. Press
Define STDs.
Step 6. Select #1, press
Label, and enter Open 3G using the character input pad that appears on the
screen.
Step 7. Select
Open as the standard type (STD Type).
Step 8. Set a necessary standard coefficient.
Step 9. Enter
1G as the minimum frequency for Min Frequency.
Step 10. Enter
3G as the maximum frequency for Max Frequency.
Step 11. Press
Return.
Step 12. In the same way, select #2, press
Label, and enter Open 6G using the character input pad
that appears on the screen.
Step 13. Select
Open as the standard type (STD Type).
Step 14. Set a necessary standard coefficient.
Step 15. Enter
3G as the minimum frequency for Min Frequency.
Step 16. Enter
6G as the maximum frequency for Max Frequency.
Step 17. Press
Return.
Table 4-6 Example of standard definition
Standard Label name Frequencies to be defined
Open Open 3G 1 GHz - 3 GHz
Open Open 6G 3 GHz - 6 GHz