Agilent Technologies ENA Series Podręcznik Użytkownika Strona 191

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 190
Chapter 4 185
Calibration
Changing the Calibration Kit Definition
4. Calibration
Redefining a calibration kit
To change the definition of a calibration kit, follow the procedure below.
Defining a new calibration kit
1. Select the calibration kit to be redefined.
2. Define the type of standard. Select one from among the OPEN, SHORT, LOAD,
delay/THRU, and arbitrary impedance standards.
3. Define the standard coefficient.
4. Designate a standard class for the standard.
5. Save the data for the calibration kit that has been redefined.
Changing the pre-defined connector between male and female (e.g. OPNE (f) to OPEN
(m))
1. Select the calibration kit to be redefined.
2. Designate a standard class for the standard.
3. Save the data for the calibration kit that has been redefined.
Redefining a calibration kit
Step 1. Press .
Step 2. Press
Cal Kit.
Step 3. Select the calibration kit to be redefined according to the list below.
NOTE If the names (labels) of calibration kits were changed prior to operation, the new names
will appear on the respective softkeys.
Softkey Function
85033E Selects the “85033E” calibration kit
85033D Selects the “85033D” calibration kit
85052D Selects the “85052D” calibration kit
85032F Selects the “85032F” calibration kit
85032B Selects the “85032B” calibration kit
85036B/E Selects the “85036B/E” calibration kit
85031B Selects the “85031B” calibration kit
85050C/D Selects the “85050C/D” calibration kit
85052C Selects the “85052C” calibration kit
85038A/F/M Selects the “85038A/F/M” calibration kit
User Selects the fourth user-defined calibration kit
Przeglądanie stron 190
1 2 ... 186 187 188 189 190 191 192 193 194 195 196 ... 774 775

Komentarze do niniejszej Instrukcji

Brak uwag