Reflection measurement:
- Press Reflection.
- Connect the Reflection standard to test port 1, and press
[ ]Reflection 1
. The reflection
measurement is performed, and an asterisk is indicated.
- Connect the Reflection standard to test port 2, and press [ ]Reflection 2. The reflection
measurement is performed, and an asterisk is indicated.
- Connect the Reflection standard to the test port 3, and press [ ]Reflection 3. The reflection
measurement is performed, and an asterisk is indicated.
Match measurement:
- Press Match.
- Connect the Match standard to test port 1 and press
[ ]Match 1
. The Match measurement is
performed, and an asterisk is indicated.
- Connect the Match standard to test port 2 and press [ ]Match 2. The Match measurement is
performed, and an asterisk is indicated.
- Connect the Match standard to test port 2 and press [ ]Match 3. The Match measurement is
performed, and an asterisk is indicated.
Line1 measurement:
- Press Line1.
- Connect the Line1 standard between test ports 1 and 2, and press [ ]Line1 1-2. The Line1
measurement is performed, and an asterisk is indicated.
- Connect the Line1 standard between test ports 1 and 3, and press [ ]Line1 1-3. The Line1
measurement is performed, and an asterisk is indicated.
Line2 measurement:
- Press Line2.
- Connect the Line2 standard between test ports 1 and 2, and then press
[ ]Line2 1-2
. The Line2
measurement is performed, and an asterisk is indicated.
- Connect the Line2 standard between test ports 1 and 2, and then press [ ]Line2 1-3. The Line2
measurement is performed, and an asterisk is indicated.
6-2) When all of the necessary calibration measurements are completed, the check marks are indicated on the
Thru/Reflection/Match/Lines keys, and the Update CalCoef key appears. Press the Update CalCoef key to
activate the measured calibration data. Confirm that the indication “
Cor”
appears on the display.
7. Finishing multiport TRL/LRM calibration
To leave TRL/LRM calibration, click on the close button [X] in the upper right hand corner of the main
control panel.