Agilent E5070B/E5071B ENA Series RF Network Analyzers
TRL/LRM Calibration
First Edition
No. 16000-95026
January 2004
Przeglądanie stron 0
1 2 3 4 5 6 ... 12 13

Podsumowanie treści

Strona 1 - TRL/LRM Calibration

Agilent E5070B/E5071B ENA Series RF Network Analyzers TRL/LRM Calibration First Edition No. 16000-95026 January

Strona 2 - Notices

Reflection measurement: - Press Reflection. - Connect the Reflection standard to test port 1, and press [ ]Reflection 1. The reflection measurement

Strona 3

Multiport TRL/LRM Calibration Operating Procedure You can make 3- and 4-port TRL/LRM calibration in the same manner as 2-port TRL/LRM calibration. 1

Strona 4

5. Defining calibration kit Define the values of the calibration standards in the same manner as 2-port TRL/LRM calibration. 6. Performing calibr

Strona 5

Reflection measurement: - Press Reflection. - Connect the Reflection standard to test port 1, and press [ ]Reflection 1. The reflection measuremen

Strona 6

Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protect

Strona 7

TRL/LRM Calibration Overview The TRL/LRM Calibration VBA Macro lets you perform the 2-, 3-, and 4-port TRL (Through-Reflect-Line), LRL (Line-Reflect-

Strona 8

To perform TRL/LRM calibration, you need to prepare calibration standards that meet the following requirements: Through (zero-length, or non-zero-len

Strona 9

Installation of TRL/LRM Calibration VBA Macro NOTE Before installing this VBA macro, make sure that your E5070B/E5071B’s firmware revision is 3.5 or

Strona 10

2-port TRL/LRM Calibration Operating Procedure 1. Setting stimulus conditions Select the ENA’s active channel to the one for which you want to perfo

Strona 11

Figure 2. TRL/LRM main control panel 4. Selecting ports In the Ports menu, select the test ports for which you want to perform calibra

Strona 12

5-2) Define each standard as follows: Reference impedance Z0 Enter the reference impedance value for the through, match and line standards. Reflecti

Strona 13

5-4) If you want to save the current calibration kit definition, press the Save key and save the definition to your desired file. Saved calibration ki