Agilent Technologies Agilent 8166A Instrukcja Obsługi Strona 58

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 274
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 57
Instrument Setup and Status IEEE-Common Commands
58 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition
command:
*ESE?
syntax: *ESE?
description: The standard Event Status Enable query *ESE? returns the contents of the Standard Event
Status Enable Mask (see *ESE for information on this register).
parameters: none
response: The bit value for the register (a 16-bit signed integer value).
example: *ESE? 21<END>
command:
*ESR?
syntax: *ESR?
description: The standard Event Status Register query *ESR? returns the contents of the Standard
Event Status Register. The register is cleared after being read.
parameters none
response The bit value for the register (a 16-bit signed integer value):
Bit Mnemonic Decimal Value
7 (MSB) Power On 128
6Not used 0
5 Command Error 32
4 Execution Error 16
3 Device Dependent Error 8
2 Query Error 4
1Not used 0
0 (LSB) Operation Complete 1
example: *ESR? 21<END>
Przeglądanie stron 57
1 2 ... 53 54 55 56 57 58 59 60 61 62 63 ... 273 274

Komentarze do niniejszej Instrukcji

Brak uwag