Agilent Technologies Agilent 8166A Instrukcja Obsługi Strona 137

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 274
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 136
Signal Generation – The SOURce Subsystem Measurement Operations & Settings
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Fifth Edition 137
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes<wsp>
<value>|MIN|MAX|DEF|0
description: Sets the number of cycles.
NOTE Cannot be set while a sweep is running.
parameters: The number of cycles is an integer value.
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum programmable value and the maximum
programmable value
0: cycles continuously.
response: none
example: wav:swe:cycl 3
affects: All tunable laser modules
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes?
syntax: [:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:CYCLes?
[<wsp>MIN|MAX|DEF]
description: Returns the number of cycles.
parameters: none
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum programmable value and the maximum
programmable value
response: The number of cycles as an integer value.
example: wav:swe:cycl? +3<END>
affects: All tunable laser modules
Przeglądanie stron 136
1 2 ... 132 133 134 135 136 137 138 139 140 141 142 ... 273 274

Komentarze do niniejszej Instrukcji

Brak uwag