6-20 Agilent B2200 Users Guide, Edition 4
VXIplug&play Driver
Driver Functions
Syntax ViStatus _VI_FUNC agb220xa_esr_Q(ViSession vi, ViChar_VI_FAR errstr[ ] );
Parameters vi Instrument handle returned from agb220xa_init( ).
errstr[ ] Response from instrument.
Bit Value Message
1 “ESR_OPC”
2 “ESR_RQL”
4 “ESR_QYE_ERROR”
8 “ESR_DEVICE_DEPENDENT_ERROR”
16 “ESR_EXECUTION_ERROR”
32 “ESR_COMMAND_ERROR
64 “ESR_URQ”
128 “ESR_PON”
OTHERS “ESR_MULTI_EVENT”
agb220xa_func
This function is used to set the channel configuration to the auto configuration mode
or the normal configuration mode.
Syntax ViStatus _VI_FUNC agb220xa_func(ViSession vi, ViInt16 channel_config);
Parameters vi Instrument handle returned from agb220xa_init( ).
channel_config Configuration mode. 0 (auto) or 1 (normal).
agb220xa_groundChanCard
This function ground-enables or disables all the output ports of the specified card.
Syntax ViStatus _VI_FUNC agb220xa_groundChanCard(ViSession vi,
ViInt16 disable_enable, ViInt16 gnd_cardno);
Parameters vi Instrument handle returned from agb220xa_init( ).
disable_enable Status of the card, ground enabled or ground disabled.
0 : sets ground enabled card.