Agilent Technologies 6051A Instrukcja Użytkownika Strona 6

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 11
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 5
4
Table 60503-1 Specifications and Supplemental Characteristics (continued)
External Analog Programming 0 to 10 V (dc or ac):
Bandwidth: 10 kHz (3 db frequency)
Accuracy: ± 3% ± 10 mA (0 to 1 A range)
± 3% ± 20 mA (0 to 10 A range)
± 0.5% ± 150 mV (0 to 240 V range)
Temperature Coefficient: 150 ppm/
°
C ± 1 mA/
°
C (current ranges)
120 ppm/
°
C ± 10 mV/
°
C (voltage range)
External Current Monitor (0 to 10 V):
Accuracy: ± 3% ± 10 mA (referenced to analog common)
Temperature Coefficient: 100 ppm/
°
C ± l mA/
°
C
External Voltage Monitor (0 to 10 V):
Accuracy: ± 0.4% ± 240 mV (referenced to analog common)
Temperature Coefficient: 70 ppm/
°
C ± 1.2 mV/
°
C
Remote Sensing: 5 Vdc maximum between sense and input binding posts
Maximum Input Levels:
Current: 10.2 A (programmable to lower limits)
Voltage: 250 V
Minimum Operating Voltage: 2 V (derated to 0 V at 0 A)
PARD (20 Hz to 10 MHz noise):
Current: 1 mA rms/10 mA p-p
Voltage: 6 mV rms
DC Isolation Voltage: ± 240 Vdc between + or - input binding post and chassis ground
Digital Inputs:
Vlo: 0.9 V maximum at Ilo = -1 mA
Vhi 3.15 V minimum (pull-up resistor on input)
Digital Outputs:
Vlo: 0.72 V maximum at Ilo = 1 mA
Vhi: 4.4 V minimum at Ilo - 20 µA
SUPPLEMENTAL CHARACTERISTICS
Programmable Slew Rate (For any given input transition, the time required will be either the total slew time or a minimum
transition time, whichever is longer. The minimum transition time increases when operating with input currents under 0.2 A
and decreases with input currents over 2 A. The following are typical values; ± 25% tolerance):