Przeglądanie stron 3
2
Table 60503-1. Specification and Supplemental Characteristics
SPECIFICATIONS
DC Input Rating:
Current: 0 to 10 A
Voltage: 3 V to 240 V (minimum dc operation from 0 to 2 V for 0 to 10 A)
Power: 250 W at 40
°
C (derated to 187 W at 55
°
C)
A. OPERATING CHARACTERISTICS B. DERATED CURRENT DETAIL
Constant Current Mode:
Ranges: 0 to 1 A; and 0 to 10 A
Accuracy: (after 30 second wait):
±
0.15% ± 10 mA (both ranges)
Resolution: 0.26 mA (1 A range); 2.6 mA (10 A range)
Regulation: 8 mA (both ranges)
Temperature Coefficient: 150 ppm/
°
C ± l mA/
°
C (both ranges)
Constant Resistance Mode:
Ranges: 0.20 to 24
; 24
to 10 k
; and 240
to 50 k
Accuracy: ± 0.8% ± 200 m
with
1 A at input (24
range);
± 0.3% ± 0.3 mS with
24 V at input (10 k and 50 k
ranges)
Resolution: 6 m
(24
range); 0.011 mS (10 k
range); 0.001 mS (50 k
range)
Regulation: 10 mV with remote sensing (24
range); 8 mA (10 k and 50 k
ranges)
Temperature Coefficient: 800 ppm/
°
C ± 10 m
/
°
C (24
range);
300ppm/
°
C ± 0.03 mS/
°
C (10 k and 50 k
ranges)
Constant Voltage Mode:
Range: 0 to 240 V
Accuracy: ± 0.12% ± 120 mV
Resolution: 64 mV
Regulation: 10 mV (remote sense); 40 mV (local sense)
Temperature Coefficient: 120 ppm/°C ± 10 mV/°C