D Single-Mode/Multimode Module Performance Tests Performance Test Form Sheets
256 E6000C Mini-OTDR Users Guide, E0302
Performance Test for the Agilent E6000C with Multimode Modules Page 5 of 6
Model Agt. __________ Module Report No. __________ Date ___________
No. Test Description
I. Dynamic Range 1300 nm Wavelength
Minimum Specification
Pulsewidth Dynamic
Range
98%
Dyn Range =
Dyn Range
98%
+ 1.9dB
E6005A E6009A Meas.
Uncertainty
10
µs
1
µs
100 ns
10 ns
dB
_______
_______
_______
_______
dB
_______
_______
_______
_______
dB
34
28
22
17
dB
n/a
23
18
12
dB
_______
_______
_______
_______
II. Event Deadzone 1300 nm Wavelength
Event Deadzone Max Spec Meas.
Uncertainty
Return Loss
35dB
Conditions:
Meas. Span. 2km
Pulsewidth 10ns
E6005A, E6009A _______ m 3 m _______ m