Chapter 4 Command Reference
Common Commands
4-50 Programming Guide
4
*PRE (Parallel Poll Enable Register)
*PRE? (Parallel Poll Enable Register Query)
COMMAND *PRE<NRf>
*PRE?
Sets or queries the value of the parallel poll enable register.
QUERY RESPONSE <NRf>
RANGE 0-255
COMMENTS The parallel poll enable register is eight bits wide and has the same bit
definitions as the status byte.
The status byte and parallel poll enable registers are anded together;
the result determines the value of *IST (TRUE or FALSE).
RELATED
FRONT-PANEL
KEYS
None
*RCL (Recall Command)
COMMAND *RCL <NRf> | <non-decimal numeric>
This command restores the state of the instrument from a copy stored in
local non-volatile memory. Before the recall occurs, the current state of
the instrument is automatically saved to register 0.
<NRF> RANGE 0 to 8
<NRF> RESOLUTION 1
RELATED
FRONT-PANEL
KEYS
Shift + Menu (RECALL > 0 through 8)