Agilent Technologies 1000A Podręcznik Użytkownika Strona 25

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 158
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 24
Getting Started 1
Agilent 1000 Series Oscilloscopes User’s Guide 25
High Frequency Compensation
For the supplied passive probes:
1 Using the BNC adapter, connect the probe to a square wave generator.
2 Set the square wave generator to a frequency of 1 MHz, an amplitude of
3 Vp- p, and an output termination of 50.
3 Press the [Auto-Scale] front panel key.
4 If waveform does not appear like the Correctly Compensated waveform
shown in Figure 5, then use a nonmetallic tool to adjust the 2 high
frequency compensation adjustments on the probe for the flattest
square wave possible.
Figure 5 High Frequency Probe Compensation
High frequency compensation adjustments
Over Compensated
Under Compensated
Correctly Compensated
Przeglądanie stron 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 157 158

Komentarze do niniejszej Instrukcji

Brak uwag