Agilent Technologies InfiniiVision 7000B Series Podręcznik Użytkownika Strona 396

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 395
14 Utilities
396 InfiniiVision Oscilloscopes User’s Guide
3 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on the rear
panel to the BNC barrel connector.
4 Press the [Utility] key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear-panel CALIBRATION switch to
PROTECTED.
User Cal Status
Pressing [Utility]&Service&User Cal Status displays the summary results of the
previous User Cal, and the status of probe calibrations for probes that can be calibrated.
Note that passive probes do not need to be calibrated, but InfiniiMax probes can be
calibrated. For more information about calibrating probes see page 78.
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
Przeglądanie stron 395
1 2 ... 391 392 393 394 395 396 397 398 399 400 401 ... 427 428

Komentarze do niniejszej Instrukcji

Brak uwag